Inputs
MI3
41
Figure 25: Adjustment of Emissivity at FTC Input (Example)
7.2
Emissivity (digital)
Function:
emissivity (digital control)
Signal:
digital low/high
Terminal:
FTC1-3, GND
The box electronics contains a table with 8 pre-installed settings for emissivity. To activate these
emissivity settings, you need to have the inputs FTC1, FTC2, and FTC3 connected. According to the
voltage level on the FTC inputs, one of the table entries will be activated.
0 = Low signal (0 V)
1 = High signal (from 5 V to V
DC
)
A non-wired input is considered as not defined!
Table entry Emissivity
(Examples)
FTC3 FTC2 FTC1
0
1
2
3
4
5
6
7
1.100
0.500
0.600
0.700
0.800
0.970
1.000
0.950
0
0
0
0
1
1
1
1
0
0
1
1
0
0
1
1
0
1
0
1
0
1
0
1
Figure 26: Digital Selection of Emissivity with FTC Inputs
The values in the table cannot be changed through the control panel.
7.3
Ambient Temperature Compensation
Function:
Ambient Temperature Compensation
Signal:
0 to 5 V
DC
Terminal:
FTC2, GND
“product 1”
“product 2”
4.0 V (ε=0.9)
1.5 V (ε=0.4)
To the FTC input
of the box
R1 = 200 Ω
R2 = 500 Ω
R3 = 300 Ω
+ 5 VDC
1.800.868.7495
Содержание MI3
Страница 2: ...1 800 868 7495 info Raytek Direct ca...
Страница 9: ...20 NOTES 119 1 800 868 7495 info Raytek Direct ca...
Страница 73: ...Accessories MI3 Rev F1 05 2013 73 Figure 66 Protective Window 1 800 868 7495 info Raytek Direct ca...
Страница 113: ...ASCII Programming MI3 Rev F1 05 2013 113 1 800 868 7495 info Raytek Direct ca...
Страница 119: ...Notes MI3 Rev F1 05 2013 119 20 Notes 1 800 868 7495 info Raytek Direct ca...