Installation
24
MI3
Figure 11: Single Head Configuration with Comm Box
To increase the number of supported sensing heads, you can use a dedicated accessory, see section
10.1.1
5.2.2
Comm Box (DIN)
The multiple sensing head configuration consists of a modular communication box provided in a DIN
rail mountable plastic enclosure for supporting 4 sensing heads simultaneously. The DIN rail
communication box provides an externally accessibly user interface. The terminal strip connectors are
used to simplify the field wiring.
Figure 12: Multiple Head Configuration with DIN Rail Comm Box
5.3
Wiring, Head Cable
The user has to install the sensor cable on the communication box. It may be shortened, if necessary,
but keep a minimal length of 20 cm (7.9 in).
Do not bend the sensing head cable tighter than a radius of 25 mm (1 in.) for the standard heads (PUR
cable) and 15 mm (0.6 in.) for the high ambient temperature heads (Teflon cable) respectively!
The total sensing head cable length for all networked sensing heads must not exceed
30 m (98 ft) for MI3 and 2x30 m (2x98 ft) for MI3M!
Head 1
(RAYMI3…)
Comm Box
(RAYMI3COMM)
Power supply,
2 analog
outputs,
3 inputs
Fieldbus
Fieldbus
or
4x analog
DIN Rail Comm Box
(RAYMI3MCOMM)
Power supply
1 alarm output,
1 trigger
input
Max. 4 Sensing Heads
(RAYMI3…)
m
ax
. 30 m
(98
ft)
m
ax
. 30 m
(98
ft)
1.800.868.7495
Содержание MI3
Страница 2: ...1 800 868 7495 info Raytek Direct ca...
Страница 9: ...20 NOTES 119 1 800 868 7495 info Raytek Direct ca...
Страница 73: ...Accessories MI3 Rev F1 05 2013 73 Figure 66 Protective Window 1 800 868 7495 info Raytek Direct ca...
Страница 113: ...ASCII Programming MI3 Rev F1 05 2013 113 1 800 868 7495 info Raytek Direct ca...
Страница 119: ...Notes MI3 Rev F1 05 2013 119 20 Notes 1 800 868 7495 info Raytek Direct ca...