ASCII Programming
110
MI3
Description
Char
Format
P B S N Legal Values
Factory default Head /
Box
Presel. Emissivity
Pointer
EP
nEP
integer
0 - 7
0
H
Emissivity Source
ES
nES
integer
√
√
I = Emissivity from Internal
(by command)
E = Emissivity from
External analog input
(0V - 5V)
D = digital
selected FTC1-3
I
H
Presel. Emissivity Value EV
nEV
float
√
√
0.1 - 1.1
H
Valley hold time
F
nF
float
√
√
√
√
0.0 - 998.9 s
(999 =
∞
)
0.0 s
H
Flicker Filter
FF
nFF
integer
0…32768
0 (LT, G5)
3000 (1M, 2M)
H
Average time
G
nG
float
√
√
√
√
0 – 999.0 s
0.0 s
H
Head Address
HA
nHA
√
1 – 9
B
Connected heads
HC
string
√
e.g. !HC – no heads
!HC1 2 3 7 8
B
Registered heads
HCR
string
√
√
e.g. !HCR – no heads
!HCR1 2 3 7 8
HCR=0
–> new registration
B
Head Status Code
HEC
nHEC
16 bit hex
√
hex value of Status Code
H
Head identification
HI
?nHI
string
√
(
√
)
e.g. !7HIMIXLTS22
set at production H
Top temperature value
of output 1
H1O
(H)
float
√
√
√
H1O=
–40.0
or H=
–40.0
500°C
B
Top temperature value
of output 2
H2O
float
√
√
√
H2O=
–40.0
500°C
B
Laser Control
HL
nHL
integer
√
√
√
0=off, 1=on, 2=flash,
3=external
0
H
Head serial number
HN
?nHN
integer
√
√
e.g. !nHN98123
set at production H
Head special
HS
?nHS string
√
√
e.g. !99HSRAY
Set at
production
H
Head Firmware Revision HV
?nHV
√
e.g. 1.01
Set in FW
H
1
setting average / peak / valley / advanced hold cancels all other hold modes
2
setting average / peak / valley / advanced hold cancels all other hold modes
3
Head Status Codes (read only):
Temperature unit
BIT0 // 0 = °C, 1 = °F
Object temperature out of range
BIT1 // 1 = out of range
Ambient temperature out of range
BIT2 // 1 = out of range
Parameter error for a command
BIT3 // 1 = error
Register write error
BIT4 // 1 = error
Self-test error
BIT5 // 1 = error
{reserved}
BIT6
Ambient temperature compensation
BIT7 // 0 = off, 1 = on
4
for MI3100 heads only
1.800.868.7495
Содержание MI3
Страница 2: ...1 800 868 7495 info Raytek Direct ca...
Страница 9: ...20 NOTES 119 1 800 868 7495 info Raytek Direct ca...
Страница 73: ...Accessories MI3 Rev F1 05 2013 73 Figure 66 Protective Window 1 800 868 7495 info Raytek Direct ca...
Страница 113: ...ASCII Programming MI3 Rev F1 05 2013 113 1 800 868 7495 info Raytek Direct ca...
Страница 119: ...Notes MI3 Rev F1 05 2013 119 20 Notes 1 800 868 7495 info Raytek Direct ca...