11.3.6 Length Correction (specific Resistance)
/
m:The Micro Junior 2 can, fully
automatic, correct the measured values
regarding the length.
To activate length correction, press the
rotary knob while “
/
m” is highlighted.
This screen will appear.
Use the rotary knob, turn clockwise or
counter-clockwise to select the desired
Unit. Once the desired Unit is reached,
press OK with the rotary knob.
•
NoCorr;
/
m;
/100
m;
/
km;
/
ft;
/100
ft;
/1000
ft
The following screen will appear.
Use the rotary knob, turn clockwise or
counter-clockwise to set the desired
Test-Length.
•
The minimum Test-Length is 1mm,
maximum 20 000 m.
By pressing the rotary knob, the setting is
confirmed.
When Length Correction is activated the
following screen will appear.
To deactivate the Length Correction go to
the
/
m Menu.
Use the rotary knob, turn clockwise or
counter-clockwise to select OFF.
By pressing the rotary knob, the setting is
confirmed.
90081-2.11 Micro Junior 2