A.6 The test set Turns off more frequently
1. As the battery charge decrease the test set will automatically turn off more frequently to
reserve the battery power.
A.7 The test set di++.+ but current is flowing
1. The test set is not reading any test voltage on the “+P” & “- P” leads.
2. The resistance value is beyond the limit of the selected range (Over Range).
A.8 Erratic or erroneous readings
Possible causes
1. Test lead damaged or not connected.
2. Poor test lead connection.
3. Test set attached to a “Live” load or high interference load.
The Micro Junior 2 is designed to be trouble free.
If problems or questions do arise please contact our service support group.
90081-2.11 Micro Junior 2
Page 42 of 44