After completion of the test the system
saves the results in memory then prompts
the operator to “Open Contact” to the
device.
Once the contacts have been removed
from the device under test it is then
detected by the system.
The system then prompts the operator to “Make Contact ” to the device again to repeat the process
of the next measurement sequence. The user can Select “Print” or "Stop" at anytime that the
system is measuring.
12.2.4 Inter… (Interval)
Selection of “Inter…” mode will measure
and store, automatically, numerous single
sequential tests at timed intervals that are
preset by the operator. The test set also
measures continuously in this mode.
In this mode, the instrument will
continuously measure multiple tests until
the operator selects and presses “Stop” or
until the number of tests that were pre-
selected have been achieved.
The instrument will automatically save test
results in a memory location that will
increment by a timed interval.
Preselected variables are preset by the
operator in the "Setup" menu .
The test system indicates and counts down
the time between test results taken and
stored.
12.2.5 Timer
The "Timer" feature is typically used with
the transformer Winding Resistance option:
WR50-A1.
This feature allows the operator to record
the time required to prepare a heated
transformer winding (or any heated device)
for cooling curve measurements.
90081-2.11 Micro Junior 2
Page 34 of 44