11.3.5 Temperature Correction
Temp: The Micro Junior 2 can, fully
automatic, correct the measured values
regarding the temperature.
To activate temperature correction, press
the rotary knob while “Temp” is highlighted.
This screen will appear. Use the rotary
knob, turn clockwise or counter-clockwise
to select the desired Material.
•
NoCorr
•
Copper Tk: 234.5°
•
Alu Tk: 225.0°
•
User : Press the rotary knob, turn
clockwise or counter-clockwise
to set the desired Temperature.
Once the desired Material is reached,
press OK with the rotary knob.
The following screen will appear.
Use the rotary knob, turn clockwise or
counter-clockwise to set the desired Ref.
Temperature T
Ref
.
Once the desired reference temperature is
set, press OK with the rotary knob.
The following screen will appear.
Connect the Optional temperature probe to
the External port, when T1 is selected.
If you dont have a Optional temperature
probe. Use the rotary knob, turn clockwise
or counter-clockwise to use EXT. Press the
rotary knob, and then turn clockwise or
counter-clockwise to set the desired
temperature you measure with your other
temperature probe.
90081-2.11 Micro Junior 2