12.3 X10A / x1A (Forward / Reverse test)
12.3.1 Single
X10 or X1 Ranges are for making measurements where the current is first positive (+) DC sourced
then negative (-) DC sourced. Basically, the current flow through the device under test is reversed
or crossed. The system measures in both positive and negative sourced modes then the resultant
resistance value is displayed. This can be very beneficial when testing in areas with high
background interference. This mode also offers the highest degree of resolution.
When the X10A range is selected, this
screen will appear.
Selection of “Single” starts a
Forward/Reverse test sequence.
Straight (+) DC Sourced
Then; Reversed (-) DC Sourced.
At the end of the test sequence the
following screen appears.
The user now has the option to:
•
“GO” :Start another test.
•
“Store” : allows to save a reading into a
memory location (Archive).
•
“Print” : allows to go to the print
screen.
•
“Mode”: Select another test mode.
•
“Setup” : allows to go to the Range
Selection screen.
90081-2.11 Micro Junior 2
Page 35 of 44