12 Testing Screen Description
12.1 Range Selection
When the Micro Junior 2 is first powered on
the following "Range Selection" screen will
appear.
Test selections are initially made by turning
the rotary knob to the desired selection of
current or resistance range then pressing
the rotary knob to begin that sequence.
The following Ranges can be selected:
The resistance Ranges are a recommendations.
Current Range
Resistance Range
10 A (F/R)
1
… 40 m
10 A
1 m
… 40 m
1 A (F/R)
1 m
… 1
1 A
10 m
… 1
0.1 A
0.1
… 10
10 mA
1
… 400
1 mA
10
... 400 k
NOTE:
NOTE:
For the descriptions of the various adjustments and options for testing, this
manual will use the 10A and x10A Ranges as examples on the following
pages.
90081-2.11 Micro Junior 2
Page 31 of 44