55M4500CCE00 REV A
GPio HDIN’s configuration
If GPio HDIN/HDOUT is set to ON in any of the above screens then an extra menu option
will appear in the main calibration menu to configure the use of the HDIN/HDOUT
connections.
Scroll to “GPio Hdins configuration” and press
“
”
to enter, the screen will show
options to set each GPio that has HDIN/HDOUT set to ON:
Select the pair of inputs required, press
“
”
to highlight the option currently set, use
“
”
or
“
”
to scroll through the available options and then press
“save”
when the
required option is displayed. Refer to the wiring drawing supplied for the necessary
settings for your system.
Warning: these settings must match the relevant wiring drawing for proper operation
of the i4500 system.
Step #3: “Calibration units” setting
1‐ In the calibration menu, scroll up or down to the item "Calibration units:".
2‐ Press
“
”
to toggle between metric and imperial.
imperial
:
use x1000lbs and feet.
metric
:
use metric tons and metres (1 metric ton = 1000 kg).