O
PTIONAL
I
TEMS
157
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
5. Right Lateral SID Measure
The Encoder Cross (A) will allow to switch between DISCRETE SID, CAN or
POTENTIOMETER.
Press LIGHT (C) to confirm the setting; the next selection will be displayed:
•
DISCRETE:
lateral SID is selected via contacts on terminals M65, M66, M68, M70, M71.
•
CAN:
Lateral SID is transmitted to the collimator via CANBus.
•
POTENTIOMETER
:
if SID is measured through potentiometer, the operator tilts the colli-
mator first and then calibrates the potentiometer.
•
M
IN
R
IGHT
P
OT
. SID
Minimum distance from a surface of image receptor to the film is to be set up.
nnn
- (value read by potentiometer).
•
M
AX
. R
IGHT
P
OT
. SID
Maximum distance from a surface of image receptor to the film is to be set up.
nnn
- (value read by potentiometer).
6. Left Lateral SID Measure
The Encoder Cross will allow to switch between DISCRETE SID, CAN or
POTENTIOMETER.
Press LIGHT (C) to confirm the setting; the next selection will be displayed:
•
DISCRETE:
lateral SID is selected via contacts on terminals M65, M66, M68, M70,
M71.
•
CAN:
Lateral SID is transmitted to the collimator via CANBus.
•
POTENTIOMETER
:
if SID is measured using potentiometer, the operator tilts the
collimator first and then calibrates the potentiometer.
Right Lateral SID
DISCRETE/CAN/POTENTIOMETER
Min Right Pot. SID: nnn
... cm
Max.Right Pot. SID: nnn
... cm
Left Lateral SID
DISCRETE/CAN/POTENTIOMETER