MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
161
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
13. Setting of SID Horizontal # 4
The Encoder Cross (A) will allow to enter the measurements from 25 min to 300 cm max.
Press LIGHT (C) to confirm the setting.
14. Setting of SID Horizontal # 5
The Encoder Cross (A) will allow to enter the measurements from 25 min to 300 cm max.
Press LIGHT (C) to confirm the setting.
15. Setting of FULL-SIZE SID # 1
The Encoder Cross (A) will allow to enter the measurements from 25 min to 300 cm max.
LIGHT (C) confirms the selection and passes to the next SID Panoramic value.
16.
Setting of FULL-SIZE SID # 2
The Encoder Cross (A) will allow to enter the measurements from 25 min to 300 cm max.
LIGHT (C) confirms the selection and passes to the next SID Panoramic value.
17. Calibration of Vertical SID Limits - minimum
The display will show indications of the minimum Vertical SID value accepted by the
collimator.
The Encoder Cross (A) will allow to enter the measurements from 50 cm min to 300 cm max.
Press LIGHT (C) to confirm the setting.
Horiz. SID Meas. # 4
xxx cm
Horiz. SID Meas. # 5
xxx cm
Panoramic SID #1
xxx cm
Panoramic SID #2
xxx cm
Vertical SID Min.
xxx cm