OPERATION OF VRAE
2 - 25
Start/stop datalogging manually
If manual datalogging mode has been selected from the
normal operation menu, toggle through the menu using the
[MODE]
key until the “
Start Datalog?”
prompt is
displayed. Pressing the
[Y/+]
key turns on datalogging.
Likewise, pressing the
[Y/+]
key while “
Stop Datalog?”
is
displayed, turns off datalogging.
The other three modes of datalogging will start and stop
automatically without any user intervention:
Automatic Datalogging
Datalogging is automatically started when the monitor is
turned on.
Scheduled Start/Stop Datalogging
Data logging starts at a specific date and time, and logs data
at the specified “data logging period.” When the “stop date
and time” is reached, the monitor stops datalogging. This
mode requires that the monitor be turned on first before the
datalogging can start.
Periodic Start/Stop Datalogging
Data logging starts daily at a specific time, and logs data at
the specified “data logging period.” When the “stop time” is
reached, the monitor stops datalogging. This mode requires
that the monitor be turned on first before the datalogging can
start.