COMPUTER INTERFACE
5 - 17
This is the place where the user can specify the STEL alarm
level (ppm), TWA alarm level (ppm), low alarm level (ppm)
and high alarm level (ppm) for Tox3 sensor.
OXY Alarm Levels (%):
This is the place where the user can specify the low alarm
level (%) and high alarm level (%) for OXY sensor.
Note:
For PGM-7840, OXY Alarm Level will be replaced
by Tox4 Alarm Levels (ppm).
LEL Alarm Levels (%):
This is the place where the user can specify the low alarm
level (%) and high alarm level (%) for LEL sensor.
VOL Alarm Levels (%):
This is the place where the user can specify the low alarm
level (%) and high alarm level (%) for VOL sensor.
Alarm mode:
If
Auto
Set
(Auto reset) is chosen, the alarm will stop as
soon as the alarm condition goes away. The alarm will
remain on if the
Latched
option is chosen. To turn off a
latched alarm, press
[Y/+]
key to acknowledge the alarm
condition.
Power On Zero:
This option allows the monitor to perform a fresh air
calibration automatically when the monitor is turned on.
Care must be taken, when using this mode, to ensure that the
monitor is in a fresh air environment when turned on.
Calibration span: