TROUBLESHOOTING
8 - 10
Problem
Possible Reasons & Solutions
“Bat”
message in
operation
Reasons:
Discharged battery
Solutions:
Recharge battery
“Voltage
too high
message” in
charge mode
Reasons:
Battery fuse blown or wrong
AC adapter
Solutions:
Check battery and adapter
Excessive
pump noise
or no inlet air
pressure
Reasons:
Leaky inlet path
Defective pump
Solutions:
Check inlet connection
Replace a new lamp
Read a small
background
value
Reasons:
There is actually a small
background gas level
Sensor zero drift
Solutions:
Do fresh air calibration
(See Section 4.4.1)