COMPUTER INTERFACE
5 - 12
Figure 5.8 General Settings Tab Page
ID:
This section sets the identification that is used throughout
the datalog process. Both editing fields,
Site
and
User,
can
hold up to eight alphanumeric letters.
Security Level:
There are three levels of security. They are used to provide
protection against unauthorized changes to monitor settings
through the programming menu of the VRAE.
Level 0
allows the user to enter the programming menu
without password checking, but does not allow changes to
alarm limits, real time clock, correction factor, or user and
site information, etc. The user can still perform
calibrations.
If
Level 1
is selected, a four-digit password is required to
enter the programming menu.