COMPUTER INTERFACE
5 - 10
Site/User ID:
This section sets the identification that is used throughout the
datalog information. The
Site ID
is an eight digit alphanumeric
field. The
User ID
is an eight digit alphanumeric field.
Security Level:
There are three levels of security to provide protection against
unauthorized changes to monitor settings. They can be changed
using the programming menu of the UltraRAE when the
instrument is in the programming mode.
Level 0 -
allows the user to enter the programming menu
without password checking, but does not allow changes to
alarm limits, real time clock, correction factor, or user and
site information, etc. The user can still perform calibrations.
Level 1 -
a four digit password is required to enter the
programming menu.
Level 2 -
the programming menu may be entered without
restriction. Any changes made in the programming menu
can be saved.
Password:
The four-digit password must be entered if level 1 is selected as
the security level or the instrument is in text mode. Entry to the
programming menu requires that this password be entered.
User Mode:
This option allows user to set the instrument to “Display” mode
or “Program” mode.
In “Display” mode, the user can view several readings in the
normal operation, but cannot enter the programming menu. In
“Program” mode, user can enter the programming mode to
calibrate the monitor or change various setup in the monitor.
Password:
The four digit password must be entered if level 1 is selected as
the security level or the instrument is in text mode. Entry to the
programming menu requires that this password be entered.