APPENDIX B
B - 2
Butadiene (Polymer) C
4
H
6
Part No.: 012-3024-005
Standard Lamp
Typical Range (ppmv)
9.8 eV
0.1 - 200
Temperature Range: 5 - 40
°
C (41 - 104
°
F)
Temp (
°
C)
5-18
18-30
30-40
Temp (
°
F)
41-64
64-86
86-104
Measure Time (sec)
180
75
50
Sample Vol. (mL)
1500
600
400
Calibration should be performed at the same temperature as the measurement.
Humidity: 0 - 95% RH.
RH
<5%
50%
80%
Correction
Factor (CF)
1.0 1.5 1.6
When calibrated from a dry gas cylinder, multiply the reading by the CF to obtain
the true value.
Color Change: None
Storage Life and Conditions: Unopened tubes can be stored for 2 years in darkness at 0 -
40
°
C (32 - 104
°
F). Open tubes may be stored for up to 8 hours in clean air without
significant loss of capacity.
Cross-sensitivity:
Substance
Test
Conc.
(ppmv)*
Apparent
Butadiene
Response
Substance
Test
Conc.
(ppmv)*
Apparent
Butadiene
Response
Acrylonitrile 100
0.0
Methane 25000**
0.0
Styrene 100
0.0
Methyl Bromide#
5
3#
Ethylbenzene 200 0.0
Propane
1000 0.0
Toluene
100 0.2
Isobutane
100 0.0
Toluene 200
2
Isobutylene
50
40
Benzene 10
0.3
n-Hexane 200
0.0
Benzene 100
7
Cyclohexane
50
0.5
*
Not necessarily the maximum allowable concentration.
** Methane above 1% by volume reduces the PID response, but has no effect on tube
capacity. Butane and higher hydrocarbons reduce tube capacity
.
# Methyl bromide can be measured using a 10.6 eV lamp and a 60 sec. sampling time at
room temperature. Adjust time at other temperatures proportionately.
Note:
Unused tubes contain no hazardous components but may adsorb toxic compounds
from the environment.