TROUBLESHOOTING
8 - 10
Problem
Possible Reasons & Solutions
Read a small
background
value
Reasons:
There is actually a small
background gas level Sensor
zero drift
Solutions:
Do fresh air calibration (see
Section 4.4.1)
Reading
jumping
around
randomly
Reasons:
Incorrect calibration gas
value.
Low sensitivity to cal gas
Solutions:
Calibrate the monitor.
Use different cal gas.
Can not turn
off monitor or
corrupted
characters in
LCD display
Reasons:
Microprocessor hang-up
Solutions:
Disconnect and re-connect
battery to reset computer.
Call authorized service
center.
Calibration
error message
Reasons:
No standard gas input, weak
PID lamp, unit zeroed with
gas present
Solutions:
Make sure standard gas flows
into monitor. Zero sensor &
recalibrate, replace lamp.