MeshGuard User’s Guide
4
WARNINGS
Read Before Operating
This manual must be carefully read by all individuals who have or will
have the responsibility of using, maintaining, or servicing this product.
The product will perform as designed only if it is used, maintained, and
serviced in accordance with the manufacturer’s instructions.
Warning:
Use only the Lithium battery or external rechargeable battery provided
by RAE Systems. This instrument has not been tested in an explosive
gas/air atmosphere having an oxygen concentration greater than 21%.
Substitution of components may impair suitability for intrinsic safety.
Replacement of batteries only in non-hazardous locations.
STATIC HAZARD:
Clean only with a damp cloth.
For safety reasons this equipment must be operated and serviced by
qualified personnel only. Read and understand instruction manual
completely before operating or servicing. Any rapid up-scale reading
followed by a declining or erratic reading may indicate a gas
concentration beyond upper scale limit, which may be hazardous.
Intrinsically safe marking:
IECEx EX ia II C
IECEx number: TSA 07.0032X
UL/CSA C1D1, group A,B,C, D T4
ATEX II 1G EEx ia IIC T4