PROGRAMMING AREARAE-Gamma
4-33
4.7.9. Change Pump Speed
There are two speed settings for the pump motor in the
AreaRAE-Gamma that can be selected from the programming
menu: low (default) and high. The “high” setting (400cc flow per
minute) should be used for long lengths of tubing or when rapid
changes in input conditions are expected. The “Low” setting
(300cc flow per minute) may be selected when the operating
conditions are slow to change. Low may also be used to
prolong pump motor life; LEL sensor life and battery run time.
1. When the “Change Pump Motor Speed?” display appears,
press [Y/+], the display shows the current pump speed
selection: “Pump Speed = Low?”
2. Press [Y/+] to accept the currently displayed pump speed
and move to the next monitor setup submenu. Press [N/-]
to change to the other speed “Pump Speed = High?”
Press [MODE] to exit this submenu and move to the next
monitor setup submenu.