OPERATION OF ACCESSORIES
3-7
3.5 Dilution Fitting
An optional dilution fitting can be installed with the remote
sampling probe or Teflon
®
tubing on the gas inlet port to dilute
gas samples. This fitting is needed when the gas sample
contains less than 15% oxygen. The combustible sensor will
not function correctly when the oxygen concentration falls
below 15%. The dilution fitting will increase the oxygen
concentration so the combustible sensor may still be used
under oxygen deficient conditions. The dilution fitting may also
be used to measure combustible, VOC or toxic gases when
concentrations exceed the upper limit of the sensor range.
To use the dilution fitting, insert the fitting between the external
filter and the remote sampling probe or Tygon tubing. Set the
dilution ratio in Program Mode so the correct gas reading
displays when the dilution fitting is used. (See Section 4.8.4
Change Dilution Ratio)
WARNING
For this application, the user must have the monitor located in
a clean atmosphere outside the confined space and use a
remote access probe or Teflon
®
tubing to measure the gas
concentration.