PROGRAMMING AREARAE-Gamma
4-18
Or if the Averaging Method is the Running Average, the
AVG will display instead of the TWA. (See Section 4.7.9
Change Averaging Method)
TOX1
VOC
Gamma
1.0
10.0
AVG
LEL
OXY
2. To modify this limit, starting from the left-most digit, use
[Y/+] or [N/-] to change the digit value and press [MODE]
momentarily to advance to the next digit. The flashing digit
will move on to the next digit to its right. Repeat this
process until all the new Alarm limits are entered. Press
and hold [MODE] for 1 second to exit data entry mode. If
there is any change to the existing value, the display
shows “Save?” Press [Y/+] to accept the new value and
move to the next submenu. Press the [N/-] to discard the
changes. To preserve the previously stored alarm limit,
press and hold [MODE] for 1 second and the monitor will
exit data entry mode and move to the next submenu.
NOTE: The gamma sensor alarm limits are always shown in
µR/h.