- 75 -
Use of Database of Target Mass Values
PROCEDURE:
o
press Quick access key
o
to open a screen with data saved
in the
Database of Target Mass
Values.
o
search and select a required
target mass record press field with
its name,
o
the value of the target mass is
automatically enabled for use in
the dosing mode.
o
the display indicates value of the
target mass preceded with MINUS
sign
o
the balance is ready to work with
enabled target mass value .
Caution:
Increasing the displayed digits
automatically replaces the workspace area
including graphic representation in the
form of a bargraph.
o
dose the product until the
indicated mass value is 0.0g
o
the workspace area provides
additional data on the weighing
process.
o
The information items displayed in
the WORKSPACE area is specific
for the dosing mode.
18.10.3. Bargraph in dosing mode
A bargraph is one of the elements of the workspace area. A criterion for enabling a
bargraph is setting its parameter to YES when declaring information items for the
dosing mode.
Bargraph colours:
−
yellow: mass below the target field
tolerance
−
green: mass within the target field
tolerance
−
red: mass above the target field
tolerance
The result is recognized as correct if it is
/- 10% tolerance of the target
mass value.