- 95 -
18.16. Mass Comparison
Mass comparison mode enables comparing mass of a standard with
mass of a tested object. The mode applies two testing methods:
−
ABA [standard – test object – standard]
−
ABBA [standard – test object – test object – standard]
Each method requires carrying out a measuring series. Number of measurements in a
measuring series is determined from 1 to 20. In practice, the most frequently used
number of repetitions is 6.
The Quick access keys displayed in the bottom bar relate to the mass comparison
mode. Set their attribute to Yes (present in the bottom bar) or No (absent).
Adjustment
Print footer
Set tare
Clear counter
Enable tare
ID definition
Print header
xxx Non-standard buttons
Principle of use and function of the quick access keys has been described in chapter
18.6. of this user manual: Parameter of working mode WEIGHING – quick access
keys.
Other quick access keys function in accordance with the description provided in
chapter 18.7.
Information items related to mass comparison mode:
o
supplementary unit
o
tare
o
gross
o
counter
o
A (result of weighing a standard)
o
B (result of weighing a tested object)
o
ID codes
o
non-standard data
o
non-standard data project
Each information item features option YES or NO determining its displaying criterion.
CAUTION:
Defining non-standard information items is described in chapter 18.8.4.
Principles for using other buttons related to the Mass comparison mode are provided
further in this user manual.
Содержание MYA.3Y
Страница 2: ...2 APRIL 2013...