- 12 -
5.1. Level control system – automatic operation – “ALERT” mode
If the level sensor detects any level change, then
the colour of the level pictogram on the display
changes into red (colour of the smallest dot in
level pictogram), and an alert message is
displayed.
In order to level the balance press “ENTER”
button in the displayed message box.
The display changes into a control panel of the
leveling function with a top view of current level
status visible on its right side, and balance
elevation on the left. The balance pictogram is
accompanied by flashing arrows indicating a
direction for turning the balance’s adjustable feet
to restore the level status. Balance pictogram
depends on its model.
The level status and arrows directions are changing immediately on level adjusting
process thus facilitating a user in restoring the level status. A red coloured dot
indicates current level status of a balance. The adjustable feet should be turned
according to the direction indicated by arrows on the control panel, so that the dot of
the level indicator moves to the center of the inner circle of the level view.
When the dot moves to the center of the inner
circle of the level pictogram, the balance is
leveled correctly. Press OK button to accept the
settings.
Exit the level adjusting function and return to
weighing mode by pressing “OK“ button in the
bottom bar of the display.
Содержание MYA.3Y
Страница 2: ...2 APRIL 2013...