- 116 -
20. TECHNICAL DATA
20.1. Balance MYA .3Y series
Adjustment
internal
Display
5, 7” colourful touch screen display
Interface
2 x USB, 2 x RS 232, Ethernet
Working temp. + 10
o
C - + 40
o
C
Power supply
feeder 100V-240V , 50-60 Hz / 13,5-16V DC
Sensitivity drift 1x10
-6
/ºCxRt
Max humidity
65 %
Min humidity
45 %
MYA 2.3Y
MYA 0,8/3.3Y
MYA 5.3Y
MYA 11.3Y
MYA 21.3Y
Max capacity
2 g
0,8/3 g
5 g
11 g
21 g
Reading interval
1μg
1μg/0,01mg
1μg
1μg
1μg
Tare range
-2 g
- 3 g
- 5 g
- 11 g
- 21 g
Repeatability
1 μg
1 μg
1,6 μg
2 μg
2 μg
Linearity
±
3 μg
±
3 μg
±
5 μg
±
6 μg
±
7 μg
Pan size
Ø16
Ø16
Ø26
Ø26
Ø26
MYA 0,8/3.3Y.F
MYA 5.3Y.F
MYA 5.3Y.F1
Max capacity
3 g
5 g
5 g
Reading interval
1/10 μg
1 μg
1 μg
Tare range
- 3 g
- 5 g
- 5 g
Repeatability
1 μg
2,1 μg
2,1 μg
Linearity
±
3 μg
±
5 μg
±
5 μg
Pan size
∅
60
∅
100
∅
160
* - in case of any problems with static electricity, it is recommended to increase ambient
humidity to 50 %. In particular cases of static electricity phenomenon, and if maintaining
humidity level of approximately 50 % is impossible, use an air ionizer for eliminating
electrostatic effect on balance indications.
Содержание MYA.3Y