- 98 -
18.16.3. Mass comparison – report from a measuring series
Individual measurements in a series and a report from a measuring series is
not printed in an automatic manner. The criterion for printing of the
information is set in a STADNARD printout – FOOTER.
PROCEDURE:
o
Press field with name of a user.
/another accessible path:
Setup
→
Users/
o
To open a screen with list of users – select
name of a user for whom the settings are to
be changed
o
Select field with pictogram PRINTOUTS and
next select field name STANDARD
PRINTOUT
o
Select field named :FOOTER and scroll
down the menu content using the side
navigating panel
o
Set REPORT PRINTOUT option to YES.
This setting triggers a printout for
measurements and a report from comparison
process.