- 94 -
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is activated by
means of other device: keypad, touch panel, proximity sensors.
Send all implemented commands
Format:
PC CR LF
Command:
PC CR LF
- send all implemented commands
Response:
PC_A_”Z,T,S,SI…”
- command carried out, the terminal displays all
implemented commands.
Give balance type
Format: BN <CR><LF>
Response options:
BN_A_”x” <CR><LF>
- command understood, response: balance type
BN_I <CR><LF>
- command understood but not accessible at this moment
x
– series of types for a particular balance (in between inverted commas), with general balance type
in front
An example:
Command:
BN <CR><LF>
– return balance type
Response:
BN_A_”AS”
– balance type: ”AS R”
Give max capacity
Format: FS <CR><LF>
Response options:
FS_A_”x” <CR><LF>
- command understood, response: Max capacity
FS_I <CR><LF>
- command understood but not accessible at this moment
x
– Max value of reading units (in between inverted commas)
An example:
Command:
FS <CR><LF>
– return Max capacity
Response:
FS_A_”220.0000”
– Max capacity: ”220 g”
Give program version
Format: RV <CR><LF>
Response options:
RV_A_”x” <CR><LF>
- command understood, response: program version
RV_I <CR><LF>
- command understood but not accessible at this moment
x
– program version (in between inverted commas)
An example:
Command:
RV <CR><LF>
– return program version
Response:
RV_A_” 1.1.1”
– program version: ”1.1.1”
Set AUTOZERO function
Format: A_n <CR><LF>
Response options:
A_OK <CR><LF>
– command carried out
A_E <CR><LF>
– error in-course of command execution, no parameter or incorrect format
A_I <CR><LF>
– command understood but not accessible at this moment
n
– parameter, decimal value determining autozero settings
n →
0 – autozero off
1 – autozero on
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
A_1<CR><LF>
– turn autozero function on
Response:
A_OK<CR><LF>
– autozero function is on
AUTOZERO function operates until it is turned off by A 0 command.
Set ambient conditions state
Format: EV_n <CR><LF>
Response options:
EV_OK <CR><LF>
– command carried out
EV_E <CR><LF>
– error in-course of command execution, no parameter or incorrect