- 96 -
FIG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, filter symbols
Example:
Command:
FIG<CR><LF>
-
give current
filter
Response:
FIG_2_OK<CR><LF>
- currently set
average filter
Set Value release
Format: ARS_n <CR><LF>
Response options:
ARS_OK <CR><LF>
– command carried out
ARS_E <CR><LF>
– error in-course of command execution, no parameter or
incorrect format
ARS_I <CR><LF>
– command understood but not accessible at this moment
n
– parameter, decimal value determining value release options
n →
1 – fast
2 – fast+reliable
3 – reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release settings
are assigned to the working mode
An example:
Command:
ARS_2<CR><LF>
– set value release parameter to
fast+reliable option
Response:
ARS_OK<CR><LF>
– fast+reliable option set
Give current value release
Command overview:
On receipt of a command, the terminal sends information on currently value release.
Format:
ARG <CR><LF>
Response options:
ARG_x_OK<CR><LF>
- command carried out, returns set value release
ARG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, value release symbol
Example:
Command:
ARG<CR><LF>
- give current value release
Response:
ARG_2_OK<CR><LF>
- currently set value release parameter to
fast+reliable option
Set last digit
Format: LDS_n <CR><LF>
Response options:
LDS_OK <CR><LF>
– command carried out
LDS_E <CR><LF>
– error in-course of command execution, no parameter or incorrect format
LDS_I <CR><LF>
– command understood but not accessible at this moment
n
– parameter, decimal value determining last digit settings
n →
1 – always
2 – never
3 – when stable
CAUTION!
The numbering is assigned to a particular last digit option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, last digit settings
are assigned to the working mode.