- 61 - 

 

An exemplary report: 

 

 

The report may be reprinted upon pressing 

  button. In order to finish, press 

  button. The 

software returns to the main window of the mode and a next measurement can be performed. The 
balance operates with previously set parameter values (liquid, temperature) thus shortening time 
needed for the exact measurement.  

11.10.  STATISTICS 

Statistics is a working mode allowing to acquire data from series of measurements and to produce 
statistics using the acquired data. Settings of this function determine which data is displayed. 
 

F Shortcut Keys 

You can declare which particular weighing function is to be launched using F1, F2, F3 or F4 keys.  
For procedure informing you how to declare the particular function, read 

F Shortcut Keys

 section. 

 

Info field provides the following information: 

 

N  (number of samples), 

 

Sum  (total weight of samples within a series) 

 

AVG (average value of the series) 

 

Min   (minimum value in a series) 

 

Max   (maximum values in a series) 

 

DIF (difference between MAX and MIN within the series)

 

 

SDV P (standard deviation of population)

 

 

SDV S (standard deviation of a sample)

 

 

RDV P (coefficient of variance for population)

 

 

RDV P (coefficient of variance for the sample)

 

 

11.10.1. Means of Operation   

 

Enter <STATISTICS> mode. 

 

Place the first load on the weighing pan. 

 

 

When the indication is stable, confirm the measurement by pressing 

  button, the 

measurement is saved in the balance memory, next it is automatically printed out with the 
measurement number. 

 

Remove the load from the weighing pan 

 

Carry out measurements for other loads within this series 

 

Содержание AS PLUS Series

Страница 1: ...R Series Balances AS R2 PLUS Analytical Balances PS R1 Precision Balances PS R2 Precision Balances PS R2 H Precision Balances USER MANUAL IMMU 03 45 03 21 EN www radwag com ...

Страница 2: ...ding this it means that you are bound to achieve success You have purchased a device that was designed and manufactured to give you years of service Congratulations and thank you for selecting RADWAG product MARCH 2021 ...

Страница 3: ...s 12 3 6 Powering the Device 14 3 7 Connecting Additional Hardware 15 3 8 Information on the Balance 15 4 Keyboard Buttons Function 15 5 Start Up 16 5 1 Temperature Stabilization Period 16 5 2 Ambient Conditions State Indication 16 5 3 User Menu 16 5 4 Logging 18 5 5 Units 19 5 6 Temporary Measuring Unit 19 5 7 Units Accessibility 19 5 8 Start Unit Selection 20 5 9 Custom Unit 20 6 Miscellaneous P...

Страница 4: ... Pan Weighing 48 11 3 Counting Parts of the Same Mass 49 11 3 1 PARTS COUNTING Mode Settings 49 11 3 2 Setting the Reference Mass Mass Determination from the Sample of Known Quantity 49 11 3 3 Setting the Reference Mass Entering Mass Value 50 11 4 Checkweighing 51 11 4 1 Declaring Threshold Values 51 11 5 Dosing 52 11 5 1 Setting Target Mass Entering Mass Value 52 11 6 Percent Weighing Control wit...

Страница 5: ... Port Settings 72 12 3 Port USB 73 13 PERIPHERAL DEVICES 76 13 1 Computer 76 13 1 1 Computer Connection Port 76 13 1 2 Continuous Transmission 77 13 1 3 Printouts Interval for Continuous Transmission 78 13 1 4 Cooperation with E2R 78 13 1 5 Printout 78 13 2 Printer 79 13 3 Barcode Reader 81 13 4 Additional Display 81 13 5 External Buttons 82 14 Cooperation with Peripherals 83 14 1 Transferred Data...

Страница 6: ... 6 1 GENERAL INFORMATION 1 1 DIMENSIONS AS PLUS series PS 200 2000 PS 1000 PS 2100 PS 10100 ...

Страница 7: ...n adverse operating conditions condensed dust drops of water falling down at different angles typical for IP 54 The PS R2 H balances are equipped with weighing pans round ø115mm or square 195x195mm Balances with pans of a smaller size feature draft shield as well Additional asset of the PS R2 H balances are their interfaces build in a hermetic closed housing which is separated from the balance The...

Страница 8: ... socket 4 plug the power supply to the mains CAUTION While cleaning it is advisable to follow below precautions Additionally it is a must to disconnect the device from mains and unplug all the peripherals printers computer etc remembering at the same time to protect ports by means of stoppers Only thus prepared device can be cleaned Upon completion of the cleaning process the device may be connect...

Страница 9: ...and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the u...

Страница 10: ...ghing pan shields inserts The device components shall be packed into an original packaging thus being protected against potential damage during transportation 3 UNPACKING AND INSTALLATION 3 1 PLACE OF USE AND ASSEMBLING The balance should be stored and used in locations free of vibrations and shakes free of air movement and dust ambient air temperature should not exceed the range of 10 C 40 C ambi...

Страница 11: ... a transport lock 1 gently press the transport lock and turn it accordingly to OPEN instruction keep the transport lock should your balance be transported in the future Install components following the above diagram bottom insert 2 centring ring embossment side up 3 draft shield 4 weighing pan 5 PS 200 2000 PS 1000 PS 2100 PS 10100 remove tape protecting the grounding spring located on one of the ...

Страница 12: ...cess if needed In the case when contamination is hard to remove e g adhesive rubber resin polyurethane foam residues etc you can use a special cleaning agents based on a mixture of aliphatic hydrocarbons that do not dissolve plastics Before using the cleanser for all surfaces we recommend carrying out tests Do not use products containing abrasive substances Cleaning glass components Select dissolv...

Страница 13: ...r wet sponge featuring large wholes this will help you to remove loose heavy dirt Do not use cleansers containning abrasive substances Next using cloth and cleanser water solution soap dishwashing liquid gently rub the cleaned surface Avoid using cleanser without water since it may result with damage of the cleaned surface please mind that large amount of water mixed with cleanser is a must Cleani...

Страница 14: ... construction through this opening which might negatively influence the balance parameters 3 6 POWERING THE DEVICE Balance can be connected to the mains only with a power adapter that comes standard with the particular model Nominal power supply of the power adapter specified on the power adapter data plate should be compatible to the power from the mains Plug the balance to the mains connect the ...

Страница 15: ... ON OFF If switched off balance components other than the display are powered and balance is in stand by mode F9 button of the computer keyboard Press to access data stored in a database user product tare F10 button of the computer keyboard Press to enter directly the active working mode settings F11 button of the computer keyboard Press to select working mode F5 button of the computer keyboard Pr...

Страница 16: ...and 8 hours for AS balances During the thermal stabilization the indications on the display panel can change It is recommended that ambient temperature changes at place of use were insignificant slow to change 5 2 AMBIENT CONDITIONS STATE INDICATION The function is intended to inform on unstable ambient conditions for a balance it is enabled only for AS R series balances The function controls dyna...

Страница 17: ...ttes calibration P3 COMMUNICATION P3 1 COM1 transmission parameters port COM 1 P3 2 COM2 transmission parameters port COM 2 P3 3 WIFI transmission parameters port WIFI P4 DEVICES P4 1 COMPUTER PC connection port P4 2 PRINTER printer connection port P4 3 BARCODE READER barcode connection port P4 4 ADD DISPLAY additional display port P4 5 EXT BUTTONS P5 PRINTOUT P5 1 CAL REPORT contents of the adjus...

Страница 18: ... balance The software enables the entry of 100 users with various access rights First Log In operation procedure Run home screen and press button next select LOG IN option operators database window opens with list of available users or press function button with LOG IN function assigned or press button enter users database and select ADMIN user Press button for confirmation wait to be asked for a ...

Страница 19: ...m the moment of its activation until its change or switching the balance off and on Each pressing of the button results with change of the measuring unit Units list Unit Denotation Verified balance Unit Denotation Verified balance gram g yes Taele Singapore tls no milligram mg yes Taele Taiwan tlt no kilogram kg yes Taele China tlc no carat ct yes Momme mom no pound lb no Grain gr no ounce oz no N...

Страница 20: ...a specified coefficient The units are displayed as u1 user unit 1 and u2 user unit 2 Procedure for declaration of coefficient for the custom unit Custom units are valid for verified balances exclusively 6 MISCELLANEOUS PARAMETERS You can set up parameters which influence balance operation These parameters are to be found in parameters group P6 OTHER Settings modification for particular parameters ...

Страница 21: ...he software registers stable indication for a specified time interval set in parameter P6 5 STAND BY MODE than the display goes out immediately The backlight activates upon change of indication no stability pictogram on the display or pressing any key on the balance keypad Blanking works also when balance menu is entered CAUTION It is not recommended to run screensaver mode in balances with d 0 01...

Страница 22: ...ng balance s settings to maintain the best possible repeatability and weighing time at workstation The main purpose of the function is the possibility of monitoring the above mentioned parameters at optional moment and saving records from the carried out tests in the form of printed reports of the tests that are automatically generated at the end of examination The test controls repeatability of p...

Страница 23: ...ive factor of indications to balance memory the said factor must be referred to a mass standard In other words balance adjustment shall be performed from time to time Adjustment should be carried out Before the beginning of weighing procedure If long breaks between following measuring series occur If temperature inside the balance changes more than 1 C or 2 C for AS series balances or 2 C for PS s...

Страница 24: ...ise system for monitoring temperature At each adjustment process the temperature is recorded The next adjustment is automatically initiated if temperature changes more than 1 C or 2 C AS PS series balances from the last saved temperature Adjustment triggered by elapsing time You can declare time intervals which are criteria for balance adjustment this option is only available for non verified bala...

Страница 25: ...d 7 3 2 External Adjustment The external adjustment for PS series balances should be carried out with an external mass standard of class F1 function unavailable for verified balances Procedure Run an external adjustment process the balance displays a command ordering to unload the weighing pan REMOVE MASS the weighing pan must be empty When the weighing pan is unloaded press button The balance det...

Страница 26: ...ENT REPORT P5 1 CAL REPORT is a group of parameters enabling user to declare data that is to be printed on an adjustment printout Variable Overview PROJECT Option enables naming the project name associated with a particular type of weighing The name may consist of 16 characters maximum CALIB TYPE Option enables printing out the type of the adjustment being carried out USER Option enables printing ...

Страница 27: ... Option enables printing out the value of VARIABLE 2 Header GLP printout Footer NET Option enables printing out net weight value in a basic unit calibration unit GLP printout TARE Option enables printing out the tare value in the current unit GLP printout GROSS Option enables printing out the gross mass value in the current unit GLP printout CURR RES Option enables printing out the current measure...

Страница 28: ...tandard printout can consist of 160 character long string 8 3 1 Inserting Text List of variables mutual for all working modes having the same values Printout of a character V Current net mass in the current unit N Current net mass in the basic unit G Current gross mass in the basic unit T Current tare mass in the basic unit D Current date M Current time I Balance number R Program number P Project ...

Страница 29: ...on type of data that is to be printed out Example 1 RADWAG DATE current measurement date TIME current measurement time PRODUCT MASS current mass indication SIGNATURE current working mode Enter printout content settings and design the printout using respective data variables and characters for text format Example 2 To crop the paper after the printout had been carried out by EPSON printer if the pr...

Страница 30: ...e given Variables are used for entering various data during the weighing process e g serial number or batch number The program allows to enter two variables Each can consist of max 32 characters In order to input a variable content the user needs to enter variable settings parameter P5 9 VARIABLE 1 or P5 10 VARIABLE 2 and enter the respective values using direction keys arrows on the balance keypa...

Страница 31: ...ls Permissions USER Access to parameters of the submenu Reading permission to change settings for parameter group Other except for Date and Time Permission to start and carry out all the weighing processes You can preview information in Databases and define universal variables ADV Access to editing parameters of the submenu Reading Working modes Communication Devices Other except for Date and Time...

Страница 32: ...n above 9 3 TARES TARES 10 different weights of the packaging The following data can be inserted for each packaging NAME 30 characters TARE packaging weight enter a value with the accuracy of a reading unit In order to add tare packaging weight enter tare database and add name of the tare along with respective data follow the procedure as in Users section 9 4 WEIGHINGS Weighings database is non ed...

Страница 33: ...ingle measurement To swap between measurements recorded in database use the arrow buttons or Pressing one of these buttons lets the user move to the next measurement either up or down the list To view the remaining data relating to a respective measurement first select the measurement in question next press button The software automatically shows the data relating to the measurement data is presen...

Страница 34: ... measurement Date of the measurement Time of the measurement Measurement result mass Tare value Person carrying out the measurement logged in user The name of the product that has been weighed The software saves the measurements in a so called loop i e when the measurement 100 001 is saved the measurement 1 is automatically deleted from the balance s memory The measurements saved in the balance s ...

Страница 35: ...RT 10 1 DATABASE ESPORT To export the database select an option EXPORT The following functions are available Export of all the databases Export of users databases Export of product databases Export of tares databases Export of weighings Export of weighings saved in ALIBI memory Export of user parameters After selecting ALL DATABASES option the balance software creates files on the flash drive The ...

Страница 36: ... Balance software allows the user to save data relating to a measurement on an external USB flash drive The operation can be performed as follows Plug a flash drive into USB port Press button to abandon automatically started IMPORT EXPORT option Set parameter P4 2 1 DEVICES PRINTER PORT to PENDRIVE option Return to the weighing mode From this moment on every single pressing of the button results w...

Страница 37: ...ng Peak Hold Pipettes Calibration In order to run a particular mode press button and select the mode from the list Once the button has been pressed the name of the first available function is shown or press to select working mode press to enter the selected working mode CAUTION Upon restart the balance is launched with the most recently operated working mode For settings of this function read late...

Страница 38: ...nt of weighed loads follow below procedures Start the balance with no load on the weighing pan permissible value of load on the weighing pan on balance start is 10 of its maximum capacity Load the weighing pan steadily avoiding shocks YES NO Place weighed loads centrally on the weighing pan YES NO Avoid side loading in particular side shocks ...

Страница 39: ...nd carrying out the following measurements 11 2 2 Balance Zeroing Zeroing is a function allowing to zero mass indication In order to zero mass indication press button Mass indication of zero value shall be displayed together with precise zero and stability markers Zeroing process is an equivalent for determining new zero point recognized by the balance as precise zero Zeroing is possible only for ...

Страница 40: ...ess button The balance returns to the weighing mode and the display indicates entered tare value with a minus sign Deleting tare Entered tare value can be deleted by pressing button on the overlay or by entering tare value of 0 000g see description above 11 2 4 Weighing Profiles To simplify weighing instrument operation there are 4 profiles in the software The profiles are selected and saved so th...

Страница 41: ...er profile you can modify all settings For other profiles Fast Fast dosing Precision Filter and Value release parameters are default and cannot be modified It enables customizing the instrument and utilizing its properties depending on your needs and expectations or on specific requirements for selected working mode e g DOSING as a result the device operation is quick and easy Filter level setting...

Страница 42: ...it compares balance indications at declared time intervals e g 1s provided that weighing pan is unloaded and display indication is close to zero If results vary less than the declared AUTOZERO range e g one division balance zeroes automatically marker of stable measurement result and precise zero marker are displayed If AUTOZERO function is enabled then each weighing process starts from precise ze...

Страница 43: ...nt stabilization automatic taring of the packaging mass proceeds Net marker appears in the upper part of the display Put product that is to be packed into the packaging The display shows net weight of the product Take off the product together with the packaging The balance cancels tare value packaging weight recorded in balance storage during the first step of the operating process after the gross...

Страница 44: ... option to start automatic printout and record of indications in Weighings database and Alibi database carried out in a cyclic manner in a specified time interval The interval is set in minutes in parameter P2 2 3 3 AUTO INT Interval range is 1 9999 min For automatic operation with interval it is necessary to specify interval value in min The picture presents interval setting value set to 2 min CA...

Страница 45: ...ation Function enables displaying additional information in the bottom line Depending on the needs you can choose the following options which are shown while working in WEIGHING mode BARGRAF option presents amount of used weighing capacity in a graphic form it covers 0 MAX range ...

Страница 46: ...h is to be displayed in the bottom line of the display You can insert any text consisting of 19 characters maximum If you want the declared non standard information is to be visible set parameter P2 1 2 to NSTD INF option 11 2 10 F Shortcut Keys Function enables assigning quick access for weighing functions which are to be accessible by pressing F1 F2 F3 or F4 keys You can choose between the follo...

Страница 47: ...TER All modes VARIABLE 1 All modes VARIABLE 2 All modes ON OFF LAST DIGIT All modes ENTER SAMPLE Parts Counting Dosing Percent Weighing DETERMINE SAMPLE Parts Counting Percent Weighing SET HI_LO Checkweighing exclusively START Animal Weighing Solids Density Liquids Density Pipettes Calibration RESULT Statistics Adding END Statistics Totalizing Adding DELETE LAST Totalizing Adding PROFILE All modes...

Страница 48: ... the indication returns to zero and pictograms 0 and are lit press button The balance returns to weighing with the accuracy of the I weighing range and II weighing range pictogram and marker of the last but one digit are blanked 11 2 12 Under Pan Weighing In standard analytical and precision balances loads can be weighed under a weighing pan Such means of operation requires placing a balance in an...

Страница 49: ...e described in section relating to WEIGHING mode This section covers only settings for PARTS COUNTING mode F Shortcut Keys You can declare which particular weighing function is to be launched using F1 F2 F3 or F4 keys For procedure informing you how to declare the particular function read F Shortcut Keys section 11 3 2 Setting the Reference Mass Mass Determination from the Sample of Known Quantity...

Страница 50: ...r INFORMATION function CAUTION Remember that The total mass of all the pieces put on the weighing pan must not be greater than the balance maximum weighing range The mass of a single piece cannot be lower than 0 1 reading unit of the balance If this condition is not met the balance displays message Single part mass too low While determining the number of pieces wait for the stability pictogram nex...

Страница 51: ...n ASSIGN THRESHOLDS is assigned the editing window is displayed and the value of MIN low threshold can be ascribed The ascribed value must be given in a current unit Use arrow buttons to enter low threshold value press button to confirm The software automatically proceeds to the editing window of MAX high threshold value which is to be given in the current unit Use arrow buttons to enter high thre...

Страница 52: ...lue within the following thresholds from 97 500g to 102 500g F Shortcut Keys You can declare which particular weighing function is to be launched using F1 F2 F3 or F4 keys For procedure informing you how to declare the particular function read F Shortcut Keys section 11 5 1 Setting Target Mass Entering Mass Value Procedure Press F button to which function ENTER SAMPLE is assigned the editing windo...

Страница 53: ... of control of deviation in percentage between weighed loads mass and specified reference mass The reference mass can be determined by means of a weighing process or it can be entered into memory by the user F Shortcut Keys You can declare which particular weighing function is to be launched using F1 F2 F3 or F4 keys For procedure informing you how to declare the particular function read F Shortcu...

Страница 54: ...e Additional settings list AVERAGING TIME Amount of time during which recorded measurement results are analysed Obtained data is used to determine average measurement result THRESHOLD It is a value expressed in mass measuring units In order to start measurement the indication value has to exceed the set threshold value AUTOSTART Determines start criterion for measurements whether they are to be in...

Страница 55: ...re After altering the settings return to the main window by pressing button repeatedly Next set the averaging time in seconds It is the time during which the balance software acquires measurements and based on these measurements the average result is determined The THRESHOLD parameter is not required for this measurement mode Select ANIMAL WEIGHING mode Place the container in which the measurement...

Страница 56: ... place the object that is being weighed in the container The balance automatically begins the process of measurement after exceeding the set mass threshold After weighing process completion the measurement result is locked and automatically printed In order to finish the measurement press button The program automatically returns to the main window of Animal Weighing mode 11 8 DENSITY OF SOLIDS Sol...

Страница 57: ...mometer holder 12 Additional sinker 6 Thermometer 13 Additional stand for a set of pans or a sinker 7 Flexible connector for sinker 14 Additional set of weighing pans for density determination of solids with density lower that density of water Installation procedure AS balances PS balances CAUTION Components of density determination kit shall be stored in a respective box Weighing pans and sinker ...

Страница 58: ...n the air top weighing pan 4 of the density kit and then weighing the same sample in liquid on the bottom weighing pan 2 of the density kit As the same sample is weighed in liquid the result of density determination is automatically indicated on a balance display In order to determine the density you need to 3 Install the density determination kit 4 Enter SOLIDS DENSITY function 5 Prepare a sample...

Страница 59: ...are returns to the main window of the mode and a next measurement can be performed The balance operates with previously set parameter values liquid temperature thus shortening time needed for the exact measurement 11 9 DENSITY OF LIQUIDS Liquids Density is a working mode enabling determination of density of any liquid In order to operate this mode an optional density kit supplementary equipment ap...

Страница 60: ...on 3 Prepare a sample 4 Initiate the process 5 Set the process following the displayed messages 6 Set volume of the sinker used for measuring 7 On entering the data the software proceeds to weighing 8 First place the sinker on the hook a sample mass determination in the air and confirm the indication when stable 9 Next take the sinker out of the hook put the beaker with liquid on the beaker s basi...

Страница 61: ...ming you how to declare the particular function read F Shortcut Keys section Info field provides the following information N number of samples Sum total weight of samples within a series AVG average value of the series Min minimum value in a series Max maximum values in a series DIF difference between MAX and MIN within the series SDV P standard deviation of population SDV S standard deviation of ...

Страница 62: ...er of saved measurements is displayed in a bottom line On pressing or the type of displayed information changes Selecting an option PRINT and pressing button initiates printing out the statistics data in a form of a report An exemplary report Statystyka N 9 Sum 455 600 g Avg 50 6222 g Min 49 939 g Max 51 380 g Dif 1 441 g Sdv P 0 39605 g Sdv S 0 38505 g Rdv P 0 78 Rdv S 0 82 ...

Страница 63: ...ngredients for one mixture F Shortcut Keys You can declare which particular weighing function is to be launched using F1 F2 F3 or F4 keys For procedure informing you how to declare the particular function read F Shortcut Keys section 11 11 1 Additional settings Apart from standard settings for this mode i e weighing mode settings additional parameter has been designed to describe operation of the ...

Страница 64: ...ng the mass of the last added ingredient you can return to the previous step of the procedure After changing the mass value it is possible for the user to add this mass to the total sum In such a case follow this procedure When all the ingredients have been weighed finish the procedure of totalising Procedure The message RESULT is displayed in a bottom line this means that the total result of all ...

Страница 65: ...an wherein the said control is performed by the balance software The threshold must be set in accordance with the needs prior to the measuring process F Shortcut Keys You can declare which particular weighing function is to be launched using F1 F2 F3 or F4 keys For procedure informing you how to declare the particular function read F Shortcut Keys section 11 12 1 Means of Operation Enter PEAK HOLD...

Страница 66: ...ions which are to be accessible by pressing F1 F2 F3 or F4 keys For procedure informing you how to declare the particular function read F Shortcut Keys section 11 13 1 Means of Operation Enter ADDING mode The bottom line shows total sum and data relating to the number of ingredients which are added to the total sum and the total mass if such information is selected in the settings for totalising m...

Страница 67: ...printout CAUTION On adding process completion only a summary is printed Masses of particular samples are printed on confirmation You can print the sum once again To do that press button Press button to exit the window You return to ADDING mode home screen and the data on carried out measurements is automatically zeroed You can print an intermediate sum of saved samples In such case follow the step...

Страница 68: ...tes During volume testing procedure the software determines accuracy and repeatability errors In the case of pipettes with adjustable volume errors for Max Min and Max volume are estimated All pipettes are tested for adherence to requirements of PN EN ISO 8655 2003 During the tests dosing repeatability and accuracy are monitored In order to ensure the highest accuracy of pipettes calibration proce...

Страница 69: ...ipette For pipettes with fixed volume please set parameter 1 for pipettes with adjustable volume set parameter 2 and 3 MEASUR NO allows to predefine number of measurements for each tested volume Number of measurements ranges from 6 to 20 AUTOMATIC TARING allows to turn on function of automatic taring of dosed portion of water after confirmation of measurement value set to YES Remember to select co...

Страница 70: ...ring the third control volume V3 for a tested pipette Enter V3 control volume Upon entering the data the computer software proceeds to pipettes calibration Follow the displayed description and complete the procedure By means of a pipette dose the first portion of water wait until the measurement is stable and press ENTER to confirm The computer software records the measurement If P2 3 7 AUTOMATING...

Страница 71: ...report are those that user entered at the beginning of the calibration procedure For fixed volume pipette the situation is similar if the user presses ENTER while displaying summary for volume V1 The computer software returns to the main window Now the user may either start new procedure for the same pipette or enter new data for a different pipette An exemplary report adjustable volume pipette 3 ...

Страница 72: ...ure Select communication port COM 1 or COM 2 Set respective values The RS 232 ports enable the following setting of transmission parameters Baud rate 2400 4800 9600 19200 38400 57600 115200 bit s Parity NONE ODD EVEN 12 2 WIFI PORT SETTINGS CAUTION The transmission parameters must be matched to the customer s local network To ensure proper communication with a computer via Wi Fi port set the compu...

Страница 73: ...work detected by the balance appears in the bottom line Use or buttons to select the demanded network and press button Message PASSWORD appears in the bottom line Use a computer keyboard connected to USB port to easily enter the case sensitive password using balance keyboard you can enter only upper case letters and digits Enter the network password and confirm it by pressing button Upon entering ...

Страница 74: ...btain printout upon pressing PRINT button once only supposing that control code 0C has been set as SUFIX for description of this function go to section 13 2 Printer USB port of type B is intended for Connecting balance to a PC In order to make connection of balance and computer possible it is necessary to install virtual COM port in a computer To carry out this procedure the user needs a respectiv...

Страница 75: ...wizard 2 Having completed driver installation connect balance to a computer use 1 8 meter long USB A B cable maximally in the case of already connected balance it is necessary to disconnect it and using USB cable connect the balance again 3 The system detects the new USB device and automatically starts searching a respective driver ...

Страница 76: ...e the one which has been formed while installing drivers 8 Start cooperation 13 PERIPHERAL DEVICES PERIPHERAL DEVICES menu is comprised within Parameters menu It is accessed by pressing key The menu features list of devices that can cooperate with the balance 13 1 COMPUTER COMPUTER submenu allows the user to select a port for connection of a computer running software which enables communication wi...

Страница 77: ...t be higher The value ranges between 0 quick data transfer and 9 slow data transfer around 10 characters per second In the case of Free Link printout keyboard emulator all languages without special characters are operated QWERTY keyboard layout and two languages that feature special characters Polish for Polish Programmers keyboard To enable support of this language go to printer settings on your ...

Страница 78: ...ansmission in calibration unit and in current unit activated by means of the balance see parameter P4 2 1 2 it is also valid for continuous transmission activated by command sent from a computer see point 15 COMMUNICATION PROTOCOL 13 1 4 Cooperation with E2R E2R is a control system it controls all weighing processes performed on a balance Upon activation of this option operations on some Databases...

Страница 79: ...EE LINK USB port type B for connecting a computer Tool for entering data it functions as a keyboard Upon proper modification of the printout and pressing declared balance button e g F1 header F3 footer ENTER printout the data is directly entered to computer programs e g Excel Word Notepad etc Select USB FREE LINK port and configure its settings in order to enable correct cooperation of the balance...

Страница 80: ...er and GLP printouts CAUTION It must be remembered that inserting paper crop command to SUFFIX parameter control code results in sending the code after each printout If the user wishes for one whole printout to consist of HEADER GLP PRINTOUT and FOOTER and to be cropped underneath the FOOTER then the paper crop command should be inserted only for FOOTER settings as a non standard printout with E v...

Страница 81: ...e enter code of the UK signs chart 1B5203 In such a case the P4 2 2 PREFIX parameter setting should be as follows PREFIX 1B742D1B5203 1250 code page and UK signs chart Control codes for exemplary code pages Control code Page or other command 1B742D code page 1250 1B7410 code page 1252 1B7430 code page 1254 1B5203 UK signs chart 1B5202 DE signs chart 1D564108 Paper crop 0C Form feed for PCL printer...

Страница 82: ...electrical outlet Connect the printer or terminal to COM 1 or to the electrical outlet determine the PRINTER BALANCE transfer perameters Every time the user presses TARE and PRINT external button the balance reacts as if TARE and PRINT buttons of the balance keyboard were pressed If you need to use only one button connect it directly to COM 2 port or use the electrical outlet In order to ensure th...

Страница 83: ... from a peripheral device see additional functions Value indicated on a display can be sent via COM port or USB port in the following form stable data is sent immediately on stabilisation of weighing result button unstable on pressing button display status is sent immediately to a peripheral device on a printout such status is marked with symbol located in front of the weighing result This option ...

Страница 84: ...upon pressing key Printout format for verified balances with one digit marked 1 2 3 4 14 15 16 18 19 20 stability marker space character mass space unit CR LF Stability marker space if stable if unstable if high limit is out of range v if low limit is out of range Character space for positive values for negative values Mass 11 characters right justification digit markers contained Unit 3 character...

Страница 85: ...d understood and in progress XX_I CR LF command understood but not accessible at this moment XX _ CR LF command understood but max threshold is exceeded XX _ v CR LF command understood but min threshold is exceeded XX _ E CR LF time limit exceeded while waiting for stable measurement result time limit is balance characteristic parameter Where XX stands for command name next 1 3 4 5 6 7 8 16 17 18 ...

Страница 86: ...ng unit SU Send stable measurement result in current measuring unit SUI Immediately send measurement result in current measuring unit C1 Switch on continuous transmission in basic measuring unit C0 Switch off continuous transmission in basic measuring unit CU1 Switch on continuous transmission in current measuring unit CU0 Switch off continuous transmission in current measuring unit DH Set min che...

Страница 87: ...ES_CR LF command not recognised XX _ E CR LF time limit exceeded while waiting for stable measurement result time limit is balance characteristic parameter Where XX name of a sent command _ spaces Commands Overview Zero balance Format Z CR LF Response options Z_A CR LF command understood and in progress Z_D CR LF command carried out Z_A CR LF command understood and in progress Z_ CR LF command und...

Страница 88: ...MASS FRAME response mass value in a basic measuring unit Response format 1 2 3 4 5 6 7 15 16 17 18 19 20 21 S space stability marker space character mass space unit CR LF An example S CR LF command send form a computer S _ A CR LF command understood and in progress S _ _ _ _ _ _ _ _ _ _ 8 5 _ g _ _ CR LF command carried out response mass value in a basic measuring unit where _ space Immediately se...

Страница 89: ... 8 2 3 7 _ k g _ CR LF command carried out immediate response mass value in a current measuring unit where _ space Switch on continuous transmission in a basic measuring unit Format C1 CR LF Response options C1_I CR LF command understood but not accessible at this moment C1_A CR LF command understood and in progress MASS FRAME response mass value in a basic measuring unit Response format 1 2 3 4 5...

Страница 90: ...d out Response format 1 2 3 4 12 13 14 15 16 17 18 19 U H space mass space unit space CR LF Mass 9 characters with right justification Unit 3 characters with left justification Set mass value of a single item only for PARTS COUNTING Format SM_XXXXX CR LF where _ space XXXXX mass format Response options SM_OK CR LF command carried out SM_I CR LF command understood but not accessible at this moment ...

Страница 91: ...cifying calibration start Enable automatic internal adjustment of the balance Format IC0 CR LF Response options IC0_I CR LF command understood but not accessible at this moment IC0_OK CR LF command carried out Operation is disabled for verified balances Give balance serial number Format NB CR LF Response options NB_A_ x CR LF command understood response serial number NB_I CR LF command understood ...

Страница 92: ...F accessible working modes are given in return mode number name OK CR LF command carried out An example 2 Command OMI CR LF return accessible working modes Response OMI CR LF 2 CR LF 4 CR LF 12 CR LF accessible working modes are given in return mode number OK CR LF command carried out Set working mode Command overview Command sets particular working mode Format OMS_n CR LF Response options OMS_OK ...

Страница 93: ...m gr ti N baht tola msg u1 u2 next CAUTION if x next the command swaps to another avaialable unit on the list it simulates button pressing An example Command US_mg CR LF set mg unit Response US_mg_OK CR LF mg set as a current unit Give current unit Command overview command returns current unit Format UG CR LF Response options UG_x_OK CR LF command carried out response current unit UG_I CR LF comma...

Страница 94: ...rted commas An example Command FS CR LF return Max capacity Response FS_A_ 220 0000 Max capacity 220 g Give program version Format RV CR LF Response options RV_A_ x CR LF command understood response program version RV_I CR LF command understood but not accessible at this moment x program version in between inverted commas An example Command RV CR LF return program version Response RV_A_ 1 1 1 prog...

Страница 95: ... ambient conditions symbols Example Command EVG CR LF give currently set ambient conditions Response EVG_0_OK CR LF currently set ambient conditions unstable Set filter Format FIS_n CR LF Response options FIS_OK CR LF command carried out FIS_E CR LF error in course of command execution no parameter or incorrect format FIS_I CR LF command understood but not accessible at this moment n parameter dec...

Страница 96: ...nse ARS_OK CR LF fast reliable option set Give current value release Command overview On receipt of a command the terminal sends information on currently value release Format ARG CR LF Response options ARG_x_OK CR LF command carried out returns set value release ARG_I CR LF command understood but not accessible at this moment x parameter value release symbol Example Command ARG CR LF give current ...

Страница 97: ...nce lower case letters and upper case letters Response options LOGIN OK CR LF command understood new operator logged in LOGIN ERRROR CR LF command understood name or password error occurrence logging operation cannot be carried out ES CR LF command not understood format error User logout Format LOGOUT CR LF Response options LOGOUT OK CR LF command understood operator logged out ES CR LF command no...

Страница 98: ... four digits marker 5 five digits marker Mass 10 characters for net mass given in adjustment unit with dot and a floating point sign Mass unit 3 characters left justification Tare 9 characters with dot right justification when floating point tare gets turned off automatically then zero value is sent Tare unit 3 characters left justification Hidden digit quantity Number specifying quantity of hidde...

Страница 99: ...re also used for measuring absorptiveness of substances e g foamed polystyrene The rack is made of mild steel Rack height is 330mm Density determination kit for solids and liquids It is applicable with balance with resolution at least 1 mg designed for density determination of solids and liquids The procedure is fully automatic i e the user only places samples on kit s weighing pans Additional dis...

Страница 100: ... 100 ...

Отзывы: