- 89 -
SU_E CR LF
- time limit exceeded while waiting for a stable measurement result
SU_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response: mass value in a current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
space
stability
marker
space
character
mass
space
unit
CR
LF
An example:
S U CR LF
- command send form a computer
S U _ A CR LF
- command understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
- command carried out, response: mass value in a
current measuring unit
where:
_
- space
Immediately send measurement result in a current measuring unit
Format:
SUI CR LF
Response options:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in a current measuring unit
Response format:
1
2
3 4
5
6
7-15
16
17
18
19
20
21
S U I
stability
marker
space
character mass
space
unit
CR LF
An example:
S U I CR LF
– command sent from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command carried out, immediate response: mass value
in a current measuring unit
where:
_
- space
Switch on continuous transmission in a basic measuring unit
Format:
C1 CR LF
Response options:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
marker
space
character
mass
space
unit
CR
LF
Switch off continuous transmission in a basic measuring unit
Format:
C0 CR LF
Response options:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out
Switch on continuous transmission in a current measuring unit
Format:
CU1 CR LF
Response options:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in a current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
marker
space
character
mass
space
unit
CR
LF
Switch off continuous transmission in a current measuring unit
Format:
CU0 CR LF