RADIANQBIO INSTRUMENT www.radian21.com
45
10-3. Protocol
1)
Standard protocol
①
Stream mode (F-40, ID Number setup value ‘000’)
(Ex. Data +1234.5 transmission)
CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 BYTE6 BYTE7 BYTE8
ASCII
S
T
,
N
T
,
+
0
HEX
53H
54H
2CH
4EH
54H
2CH
2BH
30H
CODE BYTE9 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16
ASCII
1
2
3
4
.
5
CR
LF
HEX
31H
32H
33H
34H
2EH
35H
0DH
0AH
1) BYTE1 ~ BYTE3
: Fixed Charactor (S T ,)
2) BYTE4, BYTE5
: Normal (N T), Error (E R)
3) BYTE6
: Fixed Charactor (,)
4) BYTE7 ~ BYTE14 : DATA 8 BYTE inc/-)
5) BYTE15
: CARRIAGE RETURN
6) BYTE16
: LINE FEED
②
Command mode (F-40, ID Number setup value ‘001~099’)
-
Command mode (PC -> INDICATOR)
CODE
BYTE1
BYTE2
BYTE3
BYTE4
BYTE5
ASCII
I
D
0
1
P
HEX
49H
44H
30H
31H
50H
1) BYTE1, BYTE2
: Fixed Charactor (I D)
2) BYTE3, BYTE4
: Device number (1 ~ 99)
3) BYTE5
: Command Order (P, Z , H, R)
-
Command Chart
Command
Description
ASCII
HEX
P
50H
Transmit the current value of order equipment.
Z
5AH
Operate the current value of order equipment as ZERO.
H
48H
Hold for order equipment.
R
52H
Release hold for order equipment
Содержание RI-10W
Страница 12: ...RADIANQBIO INSTRUMENT www radian21 com 11 6 Wiring Diagram RI 10W RI 50W RI 20W ...
Страница 13: ...RADIANQBIO INSTRUMENT www radian21 com 12 RI 25W RI 27W RI 30W ...
Страница 14: ...RADIANQBIO INSTRUMENT www radian21 com 13 RI 148A RI 148B Example External Input PLC Connection ...
Страница 17: ...RADIANQBIO INSTRUMENT www radian21 com 16 ...