background image

RADIANQBIO INSTRUMENT                                                                                                            www.radian21.com 

 

31 

8-4. Digital calibration (Calibration by sensor output value) 

 

1) RI-10W, RI-27W, RI-50W, RI-148A,B 

At  the  time  of  purchasing  sensor,  the  rated  capacity  (R.C)  and  rated  output  (R.O) 

declared on the calibration sheet can be used for the calibration for easier calibration.   

If  you  do  not  know  the  output  value  of  the  sensor  or  if  you  need  to  correct  the  error 

after calibration by the output value of the sensor, refer to 8-4 Actual Load Calibration. 

 

※  Zero position will change after setup is complete. 

※ 

Measuring Mode 

  : The state in which the current measurement is being displayed 

Measuring Mode 

While 

  key  is  being 

pressed, press 

  key. 

Displays  Function  Mode 

Displays Sensor Output Calibration State   

 

Displays Rated Output value Input State   

Displays the rated capacity value Input State 

Input  the  rated  capacity  value  by  using 

  and 

 

key. And then press 

  key. 

 

Calibration Completed. 

Input  the  rated  output  value  by  using 

  and 

 

key. And then, press 

  key. 

 

Содержание RI-10W

Страница 1: ...UAL RI 10W VER 214A RI 20W VER 214A RI 25W VER 214A RI 27W VER 214A RI 30W VER 214A RI 50W VER 214A RI 148A B VER 214A Address 1604 53 Gasan digital 2 ro Geumcheon gu Seoul Republic of Korea http www radian21 com E Mail info radianqbio com ...

Страница 2: ... up 18 8 2 Function mode Function Set up 19 1 RI 10W RI 20W RI 25W RI 27W RI 50W RI 148A B function set up 19 2 RI 30W function set up 20 3 How to change the REALY setting 21 8 3 Function list 22 8 4 Digital calibration Calibration by sensor output value 31 8 5 A c t u a l l o a d c a l i b r a t e 3 4 8 6 Data Back up Restore 38 8 7 Lock Set up 39 9 Product Inspection 40 10 OPTION 10 1 BCD output...

Страница 3: ...er your need 1 3 Comparison Output You can do set up the maxium and lower limit value by the keys on the front panel and print with contact point signal of the rear panel 1 4 Data Back up All the set up values will be memorized on the flash memory so the inputted data can be saved and no need to do the re setting even in case of the interruption of the electric power or power disconnection 1 5 Wat...

Страница 4: ...et it up where there is less dust and Keep it out of direct air including salt and ion Do not use when there is inflammable gas or heavy machinery and smog Ground earth terminal Make wire separately from power system wiring and noise wiring Make sure the use of 4 line sealed cable as a sensor cable Too long cable leads to measurement error due to wiring resistance around 10meters 2 2 Attention for...

Страница 5: ...ppm Measured value display 7 segment 5 Digit Height 14mm RI 148 Height 8mm Status display LED Red LED 6PCS Red LED 5PCS Key Switch 4pcs Comparison output 4 Relay 4 Relay Contact Capacity AC 250V 3A more than 100 000 Analog out DC 0 10V or 4 20mA Selective use Option Temperature Range 10 60 80 less than 80 RH no dewing Sizes 96 48 128 mm W H D 48 48 128mm Panel cutting size 91 5 44 5 mm 45 5 45 5 m...

Страница 6: ...7 segment 5 Digit Height 14mm RI 148 Height 8mm Key Switch 4 PCS Status display LED Red LED 6PCS Red LED 5PCS Comparison output 4 Relay 3 Relay Contact Capacity Contact Capacity AC250V 3A More than 100000 Analog out DC0 10V or 4 20mA DC0 10V or 4 20mA Option Temperature range for use 10 60 80 RH No dewing Sizes 96 48 128 mm W H D 48 48 128 mm W H D Panel cutting size 91 5 44 5 mm 45 5 45 5 mm Weig...

Страница 7: ...T UP Mode Once this is pressed on the Function Mode you will return to the measuring mode Measuring Mode Once this Key is pressed the lower limit setting value will be displayed and this value also can be changed SET UP Mode The location of row for the number flickering can be moved Measuring Mode Once this Key is pressed the upper limit setting value will be displayed and the value can be changed...

Страница 8: ... the Analog output will be 0V 4mA as well SET UP Mode Once this is pressed on the Function Set up Mode you will return to the measuring mode Measuring Mode Once this Key is pressed the lower limit setting value will be displayed and this value also can be changed SET UP Mode The location of row for the number flickering can be moved Measuring Mode Once this Key is pressed the upper limit setting v...

Страница 9: ...with 1 terminal AC IN Main Power Supply Wiring Terminal Ground Terminal as an independent ground connection OUT Analog DC 0 10V DC 4 20mA Output Terminal EXC Sensor Supply Voltage Connection Terminal EXC Sensor Supply Voltage Connection Terminal SIG Sensor Ouput Signal Connection Terminal SIG Sensor Output Signal Connection Terminal SHI SHIELD connection terminal of sensor GND TXD RXD RS232C SERIA...

Страница 10: ...SW2 OFF 21 FG Frame ground terminal RI 148A RI 148B Please check the location of terminal and its use While the projected button is being pressed please insert the cable into the lower hole completely As soon as you release the button the connection will be completed At the point please slightly pull the calbe and check whether the cable is come off or not The most suitable calbe is Φ 0 5 1 Linkin...

Страница 11: ... RS485 TXD TX RXD TX 16 GND OUT Analog out Terminal DC 10V or DC 4 20mA Attention 1 When wiring please pull out the power plug Main power supply 2 Please use the thick calbe for ground terminal so that you can avoid any trouble due to the impulse voltage or surge If possible please keep it as an independent ground This must be used in the area with heavy noise If the ground is made with other devi...

Страница 12: ...RADIANQBIO INSTRUMENT www radian21 com 11 6 Wiring Diagram RI 10W RI 50W RI 20W ...

Страница 13: ...RADIANQBIO INSTRUMENT www radian21 com 12 RI 25W RI 27W RI 30W ...

Страница 14: ...RADIANQBIO INSTRUMENT www radian21 com 13 RI 148A RI 148B Example External Input PLC Connection ...

Страница 15: ...xium value among the measured values Peak hold mode It is to hold and display the maximum value of direction Absolute peak hold mode It is to hold and display the maximum value of the absolute value 2 Sample hold It is to hold and display the value at the time of Hold signal input among the measured values On Off On Off On off sensor input value Display value analog output value M e a s u r e d v ...

Страница 16: ...is value RY1 OFF Measured Value RY2 setup value Hysteresis value RY2 OFF Measured Value RY3 setup value Hysteresis value RY3 OFF Measured Value RY4 setup value Hysteresis value RY4 OFF 3 Low limit mode Measured Value RY1 setup value RY1 ON Measured Value RY2 setup value RY2 ON Measured Value RY3 setup value RY3 ON Measured Value RY4 setup value RY4 ON Measured Value RY1 setup value Hysteresis valu...

Страница 17: ...RADIANQBIO INSTRUMENT www radian21 com 16 ...

Страница 18: ...easured Value Offset RY4 setup value Hysteresis value RY4 OFF 3 Low limit mode Measured Value Offset RY1 setup value RY1 ON Measured Value Offset RY2 setup value RY2 ON Measured Value Offset RY3 setup value RY3 ON Measured Value Offset RY4 setup value RY4 ON Measured Value Offset RY1 setup value Hysteresis value RY1 OFF Measured Value Offset RY2 setup value Hysteresis value RY2 OFF Measured Value ...

Страница 19: ...on mode Function mode Acess to each function setup mode Please refer 8 2 function mode Digital calibration mode It is to calibrate into the sensor s output value No need to prepare for the actual load stand weight Please refer 8 4 how to calibrate Actual load calibration mode It is to calibrate by adding the actual load standard weight or the load you know Please refer 8 5 how to calibrate Measure...

Страница 20: ...ow to set function for RI 10W RI 20W RI 25W RI 27W RI 50W RI 148A B The setting value of Function mode page 22 29 can be changed the key operation below F 00 F 52 Measuring mode While key is being Press press key Change to and key Select and save by key Completion ...

Страница 21: ...ian21 com 20 2 How to set function for RI 30W The setting value of Function mode page 22 29 can be changed the key operation below F 00 F 52 Measuring mode While key is being Press press key Change to and key Select and save by key ...

Страница 22: ...t key RY1 or key RY3 to display the currently set value change it to the desired value and press the key to save it Limit mode F 20 comparison mode 1 2 For RI 148A B models you can change the settings of the three Relays RY1 RY2 RY3 Measuring mode Press key Input value When set the value the key operation is and key refer to page6 Front Panel ...

Страница 23: ...king time 0 0 5 0 sec 0 0 F 12 Auto zero at start 0 1 0 F 13 Base offset 19999 99999 0 F 20 Comparison mode Decision High limit Low limit Low High 0 Decision F 21 Hysteresis 0 99 0 F 30 DAC mode b_05v b_10v b_02A b_42A U_05v U_10v U_02A U_42A b_10v F 31 DAC capacity 0 99999 30000 F 32 DAC speed 0 1 0 F 33 Zero adjustment 999 999 000 F 34 Span adjustment 999 999 000 F 40 ID Number 000 255 000 F 41 ...

Страница 24: ...ay filter F 03 Hold mode F 04 BCD Busy time X F 10 Auto zero tracking X F 11 Auto zero tracking time X F 12 Auto zero at start X F 13 Base offset X F 20 Comparison mode F 21 Hysteresis F 30 DAC mode F 31 DAC capacity F 32 DAC speed X F 33 Zero adjustment F 34 Span adjustment F 40 ID Number F 41 Baud rate F 42 Protocol F 50 Display reverse mode X F 51 Unit of force X X F 52 Key disabling ...

Страница 25: ...alue 60 Display data Setting 0 360 It sets Pulse 1 circle of Rotary body detection sensor RI 30W can set up Pulse number from 1 to 360 F 01 Division Minimum display unit setup Standard setup value 1 Display data Setting 1 Displayed In 1 0 1 2 3 4 2 Displayed In 2 0 2 4 6 8 5 Displayed In 5 0 5 10 15 10 Displayed In 10 0 10 20 30 20 Displayed In 20 0 20 40 60 50 Displayed In 50 0 50 100 150 F 02 Di...

Страница 26: ...e RI 10W RI 20W RI 25W RI 27W RI 148A B 200Hz RI 50W 1kHz F 04 BCD Busy time BCD Output time setup Standard setup value 0 100 Display data Setting 0 050 50 ms 0 100 100 ms 0 200 200 ms 0 500 500 ms 1 000 1000 ms F 10 Auto zero tracking Auto zero operation range setup Standard setup value 00 Display data Setting 00 00 Auto zero is not used 99 01 99 Set up the operation range of auto zero If the ind...

Страница 27: ...erence value the comparative output operation is operated by sum or difference based on the set value see 7 2 How to use comparison ouput function 99999 F 20 Comparision mode Comparision output mode setup Standard setup value 0 Display data Setting 0 Decision mode Relay RY1 Low RY2 OK RY3 High Output 1 High limit mode Relay RY1 RY2 RY3 RY4 Output 2 Low limit mode Relay RY1 RY2 RY3 RY4 Output 3 Low...

Страница 28: ... b_10V Display data Setting b_05v 0 5V output unidirectional unipolar b_10v 0 10V output b_02A 0 20mA output b_42A 4 20mA output U_05v 0 5V output bidirectional unipolar U_10v 0 10V output U_02A 0 20mA output U_42A 4 20mA output F 31 DAC capacity Analog Outupt value setup Standard setup value 30000 Display data Setting 0 Set the rated capacity of the Analog output If Setting value 10000 Display da...

Страница 29: ... F 13 Base offset or the F 50 Display reverse mode setting NOTE RI 10W RI 20W RI 25W RI 27W RI 148A B 200Hz RI 50W 1kHz F 33 Zero adjustment analog output zero adjustment Standard setup value 000 F 34 Span adjustment analog output range adjustment Standard setup value 000 F 40 ID Number Communication Device Number setup Standard setup value 000 Display data Setting 000 000 Device number is not set...

Страница 30: ...up Standard setup value 0 Display data Setting 0 Standard protocol 1 Modbus RTU Protocol 2 Add unit 3 Ver6 4 protocol F 43 Interval adjustment Adjusting packet transmission interval Standard setup value 0 015 Display data Setting 0 000 Not setup 0 010 about 10ms 0 015 about 15ms 0 020 about 20ms 0 025 about 25ms 0 030 about 30ms F 50 Display reverse mode Direction setting mark reverse Standard set...

Страница 31: ...n modes of 8 3 and 8 4 are not available After calibrating in kgf set value 0 set the force unit as desired Note2 If F 51 is set to MPa set value 4 the F 00 Decimal point setting cannot be changed F 52 Key disabling Front key locking set up Standard setup value 0000 Display data Setting 0 0 0 0 Hold key Lock 1 Release 0 Hi key Lock 1 Release 0 Lo key Lock 1 Release 0 Zero key Lock 1 Release 0 Set ...

Страница 32: ...e error after calibration by the output value of the sensor refer to 8 4 Actual Load Calibration Zero position will change after setup is complete Measuring Mode The state in which the current measurement is being displayed Measuring Mode While key is being pressed press key Displays Function Mode Displays Sensor Output Calibration State Displays Rated Output value Input State Displays the rated c...

Страница 33: ...you need to correct the error Zero position will change after setup is complete Measuring mode While key is being Press press key Display Function mode Display Sensor output value calibration mode Display output type select mode Display R C value input mode Calibration end Select the sensor output type by using and key And then Press key DC0 10V resistive displacement sensors DC4 20mA DC0 5V DC1 5...

Страница 34: ...8 5 Actual load calibration if you need to correct the error Zero position will change after setup is complete Measuring Mode The state in which the current measurement is being displayed Measuring mode While key is being Press press key Display Function mode Display Sensor output value calibration mode Display output type select mode Display R C value Rated capacity input mode Input the rated cap...

Страница 35: ...will be deleted c Change the current value to Zero 0 with key or external Zero input without loading the sensor After loading with a prepared standard load change the value displayed in the indicator to the value of the standard load by referring to the setup sequence chart below Measuring mode While key is being Press press key Display Function mode Display Sensor output value calibration mode Di...

Страница 36: ...he sensor can be increased In the actual load calibration mode calibration is possible at the forward output of the sensor and calibration is not possible at the reverse output of the sensor If an incorrect value is entered during setup re calibration with 8 4 Digital calibration delete all data of the values changed to the actual load calibration setting Then re run the actual load calibration se...

Страница 37: ...rd load of about half or more than half of the sensor capacity b The value shown in the indicator sensor capacity standard load rated output value R O c 8 4 Calibrate with Digital Calibration ex 1 Assuming that the sensor capacity is 10 kgf and that the value expressed by applying the standard load of 5 kgf is 10013 10013 10 5 20026 Rated output 2 0026 mV V ex 2 Assuming that the sensor capacity i...

Страница 38: ...n and calibrate the rated output value R O And the user can also calibrate with 8 5 Actual Load Calibration 4 Notes for calibrating RI 20W When calibrating with 8 4 Digital Calibration the error can be reduced by modifying the rated capacity value displacement Please correct it using the following method a Set calibration with 8 4 Digital Calibration to delete previous data b At this point select ...

Страница 39: ...urrent set up state RESTORE Restore the current set up state Measuring mode While key is being pressed Press key After 1 sec for display Display Please select YES or NO by using key YES SAVE DATA NO Do not save DATA 1 DATA BACK UP Power is off While key is being pressed turn the power on After 1 sec for display Display Please select YES or NO by using key YES Restore DATA NO Do not restore DATA 2 ...

Страница 40: ...mended to set the Lock At the first stage please start while the power is OFF Related Function when Lock is set Function related to calibration Power is OFF While key is being pressed please turn the power ON Please keep pressing key while this is being displayed or display Set up is completed Please select lock or release by pressing key ON Lock is selected Off Lock release is selected ...

Страница 41: ... load calibration at plus If the number on the Display does not match Out of maximum display range 19999 99999 Re check the 8 4 Digital Calibration setting and the F 51 Force unit setting Restore to the settings saved by RESTORE execution Pressing key in FUNC state does not move to calibration mode Locked State at 8 7 Lock Set up State in which units of conversion at F 51 Force unit is established...

Страница 42: ...slightly with rotation RI 30W Faulty cable connection Faulty detection of proximity sensors Error setting F 00 Pulse Gear Count page 24 Prove proximity sensors operation Lamp Check the distance between the proximity sensor and the gear approximately 1mm Confirm F 00 Pulse Gear Count setting page 24 ...

Страница 43: ...gic Control Computer PIN Arrangement PIN No SIGNAL PIN No SIGNAL 1 GND 20 Negative Polarity 2 1 100 21 BUSY 3 2 100 22 RLY1 4 4 100 23 RLY2 5 8 100 24 RLY3 6 1 101 25 RLY4 7 2 101 26 NC 8 4 101 27 NC 9 8 101 28 IN1 HOLD on off 10 1 102 29 IN2 ZERO 11 2 102 30 IN3 NC 12 4 102 31 IN4 NC 13 8 102 32 NC 14 1 103 33 NC 15 2 103 34 NC 16 4 103 35 Internal test voltage 5V 17 8 103 36 NC 18 1 104 37 Exter...

Страница 44: ...ut TIME T can set to Function F 04 BCD Mode 50 100 200 500 1000 Signal logic BCD DATA output Negative logic Negative BUSY output DATA READ L Polarity output H L RALAY output RY1 RY4 L BCD output circuit INDICATOR External device External VCC 50V Max Current 500mA Max ...

Страница 45: ...tion 02 RS232C Indicator Host PC TXD Transmission Data RXD Receive Datea No 2 Pin RXD Receive Datea TXD Transmission Data No 3 Pin COM Ground GND Ground NO 5 Pin Option 03 RS485 2 Port Port Type EIA 232C EIA 485 Method Full duplex asynchronous method Half duplex asynchronous method Baud rate Select one of 2400 4800 9600 19200 38400 57600 bps Parity No parity Data bit 8 bit Stop bit 1 bit external ...

Страница 46: ...BYTE4 BYTE5 Normal N T Error E R 3 BYTE6 Fixed Charactor 4 BYTE7 BYTE14 DATA 8 BYTE including 5 BYTE15 CARRIAGE RETURN 6 BYTE16 LINE FEED Command mode F 40 ID Number setup value 001 099 Command mode PC INDICATOR CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 ASCII I D 0 1 P HEX 49H 44H 30H 31H 50H 1 BYTE1 BYTE2 Fixed Charactor I D 2 BYTE3 BYTE4 Device number 1 99 3 BYTE5 Command Order P Z H R Command Chart Co...

Страница 47: ...ATA 8byte Including 5 BYTE15 CARRIAGE RETURN 6 BYTE16 LINE FEED Stream mode F 42 Protocol setup value 2 Ex Data 1234 5 transmission CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 BYTE6 BYTE7 BYTE8 BYTE9 ASCII S T N T 0 1 HEX 53H 54H 2CH 4EH 54H 2CH 2BH 30H 31H CODE BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16 BYTE17 BYTE18 ASCII 2 3 4 5 k g CR LF HEX 32H 33H 34H 2EH 35H 6bH 67H 0DH 0AH 1 BYTE1 BYTE3 Fixed...

Страница 48: ... 2 BYTE3 BYTE6 Fixed Charactor N T 3 BYTE7 BYTE14 DATA 8 BYTE Including 4 BYTE15 CARRIAGE RETURN 5 BYTE16 LINE FEED Command mode F 40 ID Number setup value 001 099 Command mode PC INDICATOR CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 ASCII I D 0 1 P HEX 49H 44H 30H 31H 50H 1 BYTE1 BYTE2 Fixed Charactor I D 2 BYTE3 BYTE4 Device number 1 99 3 BYTE5 Command Order P Z H R Command Chart Command Description ASCI...

Страница 49: ...H 44H 30H 30H 31H 2CH 2BH 30H CODE BYTE9 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16 ASCII 1 2 3 4 5 CR LF HEX 31H 32H 33H 34H 2EH 35H 0DH 0AH 1 BYTE1 BYTE3 Fixed Charactor I D 0 2 BYTE4 BYTE5 Device number 1 99 3 BYTE6 Fixed Charactor 4 BYTE7 BYTE14 DATA 8byte Including 5 BYTE15 CARRIAGE RETURN 6 BYTE16 LINE FEED ...

Страница 50: ...IT16 0008 0 04 F 13 Base offset 19999 99999 R O INT32 0009 0010 0 04 F 20 Comparison Mode 0 3 R O UNIT16 0011 0 04 F 21 Hysteresis 0 99 R O UNIT16 0012 0 04 F 30 DAC mode 0 7 R O UNIT16 0013 0 04 F 31 DAC capacity 0 99999 R O INT32 0014 0015 0 04 F 32 DAC Speed 0 1 R O UNIT16 0016 0 04 F 40 ID Number 0 255 R O UNIT16 0017 0 04 F 41 Baudrate 0 6 R O UNIT16 0018 0 04 F 42 Protocol 0 1 R O UNIT16 001...

Страница 51: ...y Setpoint 4 19999 99999 R W INT32 4008 0 10 4009 0 03 Relay Setpoint 5 Option 4010 0 10 4011 0 03 0 10 Relay Setpoint 6 4012 4013 0 03 0 10 Relay Setpoint 7 4014 4015 0 03 0 10 Relay Setpoint 8 4016 Fuction code Command Name Device Address 03 0x03 Read Holding Regigters 4001 0x0FA1 4016 0x0FB0 04 0x04 Read Input Regigters 0000 0x0000 0032 0x0020 06 0x06 Preset Single Regigter 4000 0x0FA0 16 0x10 ...

Страница 52: ...0007 0 04 F 12 Auto Zero at Start X 0008 0 04 F 13 Base offset X 0009 0010 0 04 F 20 Comparison Mode 0011 0 04 F 21 Hysteresis 0012 0 04 F 30 DAC mode 0013 0 04 F 31 DAC capacity 0014 0015 0 04 F 32 DAC Speed X 0016 0 04 F 40 ID Number 0017 0 04 F 41 Baudrate 0018 0 04 F 42 Protocol 0019 0 04 F 50 Display reverse mode X 0020 0 04 F 51 Unit of Force X X 0021 0 04 F 52 Key Disabling Measurement 0030...

Страница 53: ...put command Zero 0x0001 X Hold set 0x0002 Hold reset 0x0004 4001 0 03 Relay Setpoint 1 4002 0 10 4003 0 03 Relay Setpoint 2 4004 0 10 4005 0 03 Relay Setpoint 3 4006 0 10 4007 0 03 Relay Setpoint 4 X 4008 0 10 4009 0 03 Relay Setpoint 5 Option X 4010 0 10 4011 0 03 0 10 Relay Setpoint 6 4012 4013 0 03 0 10 Relay Setpoint 7 4014 4015 0 03 0 10 Relay Setpoint 8 4016 ...

Отзывы: