31
8-4. Digital calibration (Calibration by sensor output value)
1) RI-10W, RI-27W, RI-50W, RI-148A,B
At the time of purchasing sensor, the rated capacity (R.C) and rated output (R.O)
declared on the calibration sheet can be used for the calibration for easier calibration.
If you do not know the output value of the sensor or if you need to correct the error
after calibration by the output value of the sensor, refer to 8-4 Actual Load Calibration.
※ Zero position will change after setup is complete.
※
Measuring Mode
: The state in which the current measurement is being displayed
Measuring Mode
While
key is being
pressed, press
key.
Displays Function Mode
Displays Sensor Output Calibration State
Displays Rated Output value Input State
Displays the rated capacity value Input State
Input the rated capacity value by using
and
key. And then press
key.
Calibration Completed.
Input the rated output value by using
and
key. And then, press
key.