RADIANQBIO INSTRUMENT www.radian21.com
26
F-12. Auto zero during operation (Setting Zero when operationg)
(Standard setup value : 0)
Display data
Setting
0
Not used
1
Autozero only once after power-on
F-13. Base offset (Setting
Base
value
)
(Standard setup value : 00000)
Display data
Setting
-19999
00000 : Base offset is not used
~
-19999
~
+99999
: If set to Base value and is a Zero signal , Base value is
displayed. (When Base value 10.00 , the display is 10.00 if
you press ZERO button)
When setting the reference value, the comparative output
operation is operated by sum or difference based on the set
value
.(see 7-2.How to use comparison ouput function)
+99999
F-20. Comparision mode (Comparision output mode setup)
(Standard setup value : 0)
Display data
Setting
0
Decision mode
: Relay RY1(Low), RY2(OK), RY3(High)
Output
1
High limit mode
: Relay RY1, RY2, RY3, RY4 Output
2
Low limit mode
: Relay RY1, RY2, RY3, RY4 Output
3
Low & High limit mode
: Relay RY1, RY2, RY3, RY4 Output
F-21. Hysteresis : (Refer to 7-2 How to use comparison ouput function)
(Standard setup value : 00)
Display data
Setting
00
00
: Hysteresis – not used
~
99
01 ~99 : Hysteresis – used (Decision mode is not applied)
(see 7-2.How to use comparison ouput function)
Содержание RI-10W