RADIANQBIO INSTRUMENT www.radian21.com
14
7. Components & Function
7-1. How to use Hold mode
Hold mode is largely divided into Peak Hold and Sample Hold. Please select Peak or
Sample. To input Hold, you can use Hold key on the front panel or external input. For
the operation method, please refer to the drawing below.
1) Peak Hold
: To Hold the maxium value among the measured values.
Peak hold mode : It is to hold and display the maximum value of (+) direction.
Absolute peak hold mode : It is to hold and display the maximum value of the
absolute value (+/-).
2) Sample hold
: It is to hold and display the value at the time of Hold signal input
among the measured values.
On
Off
On
Off
On
off
①
−
:
sensor input value
②
−
: Display value &
analog output value
M
e
a
s
u
r
e
d
v
a
l
u
e
time
①
②
M
e
a
s
u
r
e
d
v
a
l
u
e
time
①
②
<Peak HOLD> <Sample HOLD>
Front panel HOLD LED
External HOLD input termanal
Front panel HOLD Key