RADIANQBIO INSTRUMENT www.radian21.com
41
When the measured
value does not
increase or
increases slightly
with rotation. (RI-
30W)
• Faulty cable connection
• Faulty detection of
proximity sensors
• Error setting F-00 Pulse
(Gear) Count (page 24)
• Prove proximity sensors operation (Lamp)
• Check the distance between the proximity
sensor and the gear (approximately 1mm)
• Confirm F-00 Pulse (Gear) Count setting
(page 24)