F-51. Force unit (Conversion unit set up)
(Standard setup value : 0)
Display data
Setting
0
kgf
kg/㎠
1
N
kgf × 9.8
-
-
2
ℓb
kgf × 2.2
-
-
3
-
-
Bar
kg/㎠ × 0.98
4
-
-
MPa
kg/㎠ × 0.098
Note1) When setting the conversion unit (set value: 1 to 4), calibration modes of 8-3
and 8-4 are not available.
After calibrating in kgf (set value:0), set the force unit as desired.
Note2) If F-51 is set to MPa (set value:4), the F-00 Decimal point setting cannot be
changed.
F-52. Key disabling (Front key locking set up)
(Standard setup value: 0000)
Display data
Setting
0 0 0 0
Hold key Lock (1), Release (0)
Hi key Lock (1), Release (0)
Lo key Lock (1), Release (0)
Zero key Lock (1), Release (0)
<Set-up example>
1001 : Zero & Hold key Lock, Hi and Lo key Release