Quidel Triage MeterPro Скачать руководство пользователя страница 74

Page

74

of

86

A

Patient

ID

is

required

for

all

patient

tests

performed.

All

QC

sample

and

patient

results

are

stored

in

the

meter’s

memory

and

can

be

downloaded

to

a

laboratory

information

system

(LIS)

or

an

external

data

management

software

program

for

further

viewing

and

printing

of

quality

control

reports.

The

Supervisor

can

select

the

QC

Frequency

for

performing

external

liquid

control

samples.

When

it

is

time

to

perform

external

QC

samples,

the

user

must

test

the

external

liquid

controls

and

obtain

acceptable

results

before

patient

samples

can

be

tested.

When

QC

samples

have

been

successfully

performed,

the

user

is

allowed

to

perform

tests

on

patient

samples.

If

QC

samples

are

not

within

an

acceptable

range,

a

QC

Lockout

function

prevents

testing

of

patient

samples.

If

desired,

the

supervisor

in

a

central

lab

setting

can

bypass

this

function.

If

test

device

controls

are

out

of

range,

a

QC

Lockout

function

is

activated

and

patient

test

results

are

blocked

from

viewing

and

printing.

This

prevents

unacceptable

results

to

be

communicated

to

and

used

by

the

physician.

Additionally,

if

test

devices

are

expired,

a

QC

Lockout

function

is

activated

and

prevents

testing

using

expired

test

devices.

Quality

Control

Samples

It

is

still

valuable

to

apply

the

traditional

approach

to

quality

control

by

testing

quality

control

samples.

These

controls

will

check

the

total

integrity

of

the

system.

The

interval

for

analyzing

these

controls,

however,

can

be

extended

due

to

the

many

other

Total

Quality

Assurance

features

inherent

in

the

Quidel

Triage

tests.

The

Quidel

Triage

tests

have

been

designed

to

maximize

Total

Quality

Assurance

in

any

testing

environment.

The

combination

of

the

QC

features

of

the

Quidel

Triage

tests

reduce

the

impact

of

procedural

errors,

ensure

reagent

integrity,

and

assure

that

patient

results

are

accurate

each

and

every

time

a

test

is

performed.

Based

on

the

current

CLIA

guidelines

and

other

regulating

bodies,

Quidel

makes

the

following

QC

recommendations:

„

Run

two

levels

or

POS

and

NEG

external

liquid

control

samples

as

appropriate

with

each

new

lot

of

reagents

and

once

every

thirty

days

with

continued

use

of

the

same

reagent

lot

number

„

Run

the

QC

Device

daily.

Notes

Regarding

Software

Controls

1.

User

ID

Access

„

To

prevent

untrained

persons

from

performing

a

test,

a

valid

User

ID

(1

Ͳ

16

characters

in

length)

must

be

entered

before

access

is

given

to

the

RUN

TEST

function

if

the

User

ID

bypass

is

OFF

.

If

a

User

ID

Bypass

is

ON

,

then

anyone

can

analyze

a

test.

„

To

prevent

unauthorized

persons

from

accessing

patient

information,

a

valid

User

ID

must

be

entered

before

access

is

given

to

the

RECALL

PATIENT

RESULTS

functions

if

the

User

ID

bypass

is

OFF

.

If

a

User

ID

Bypass

is

ON

,

then

anyone

can

recall

patient

data.

RECALL

PATIENT

RESULTS

functions

include

recalling

patient

results,

printing

patient

results,

and

recalling

the

last

test

run

on

the

meter

when

the

last

test

is

a

patient

sample,

„

The

User

ID

Bypass

is

controlled

by

the

SET

PARAMETERS

function,

which

requires

a

Supervisor

CODE

CHIP

module

for

access.

„

Access

is

available

to

anyone

wanting

to

RECALL

NON

Ͳ

PATIENT

RESULTS

from

memory

or

INSTALL

CODE

CHIP.

2.

Out

Ͳ

of

Ͳ

Range

Flags

Содержание Triage MeterPro

Страница 1: ...Page 1 of 86 User Manual...

Страница 2: ...lation 15 Unpack Meter 15 Power Meter On 15 Install Paper 16 Insert Supervisor Code Chip Module 17 To Change Meter Settings 17 Change Clock 19 Set User ID 20 Change Reference Ranges Thresholds for Tes...

Страница 3: ...Purpose 49 Function 49 Last Record 49 Patient Results 50 Patient ID 50 Test Device Lot Number L N 51 User ID 52 From Date 52 Till Date 53 QC Results 54 QC Sample Lot Number L N 55 Device Lot Number L...

Страница 4: ...ning 70 Check Battery Level 70 Battery Replacement 70 Total Quality Assurance 71 Introduction 71 Quality Assurance Functions 72 Internal Device Quality Assurance 72 Meter Quality Assurance 73 Software...

Страница 5: ...ebsite quidel com If outside the U S contact local distributor or one of the Technical Support centers listed below You may also contact us at quidel com Region Phone E mail Address Europe and Middle...

Страница 6: ...er ID Ranges Clock Meter Settings Test Settings QC Parameters Bypass x New User ID x Update User ID x Delete User ID x User ID List x QC Device Freq x QC Sample Freq x Number of Controls x Num Quant S...

Страница 7: ...urer In vitro diagnostic medical device Consult instructions for use Authorized representative in the European Community Temperature limitation Humidity limitation Waste electrical and electronic equi...

Страница 8: ...Control Calibration Verification Prescription use only Contents contains Expected values Biological risks Thaw Transfer pipette Mean Standard deviation CODE CHIP module Printer Paper Peel open here A...

Страница 9: ...n use Do NOT write or place a label anywhere on the QC device as it may interfere with the device functionality Do not move the Quidel Triage MeterPro while a test is in progress Ensure all sample flu...

Страница 10: ...Installation Guides the user through first time setup Section 4 Operation Guides the user through running tests recalling results and deleting results Section 5 Service and Maintenance Procedures Pap...

Страница 11: ...d Quality Control QC functions that allow the person operating the meter to monitor quality control Unpacking The Quidel Triage MeterPro and the items that come with it are provided in a single box Qu...

Страница 12: ...e and CODE CHIP module in the QC Device box AC DC Power Converter AA Batteries Rolls of Printer Paper additional rolls are supplied in each box of test devices Supervisor CODE CHIP module and CODE CHI...

Страница 13: ...Page 13 of 86 Parts of the Quidel Triage MeterPro...

Страница 14: ...result on paper Data Port Connection point for the optional Bar Code Scanner allowing the operator to scan patient or user identification data instead of manually entering numbers or letters on the k...

Страница 15: ...When the test is completed the screen meter display will come to rest at the main menu Low Power Indicator 8 23AM 01 20 18 Time and date Menu items User Instructions RUN TEST RECALL RESULTS INSTALL NE...

Страница 16: ...details about Installing Paper see the Care and Maintenance section b Tear or cut a clean straight edge to feed into the printer Do not cut paper at an angle as the printer must sense the edge of the...

Страница 17: ...t The port is located on the left side of the meter towards the bottom front corner Note the orientation of the CODE CHIP module in the drawing below d Once the Supervisor CODE CHIP module is installe...

Страница 18: ...ACTERS MISC ID Min 1 Max 20 1 20 1 20 AUX ID ENABLE DISABLE Disabled Disabled Enabled For reference record your settings in the space provided e Press the key to save changes to ID Settings Display Se...

Страница 19: ...and press the key b Select CLOCK using the keys and press the key c Select the HOUR MINUTE area using the keys Type in the correct time using the number keys d Press the key to move to AM or PM e Sele...

Страница 20: ...e keys Type in the new value for the HIGH or LOW cutoff as appropriate e Press the key to save changes For more details about Ranges Settings see page 33 Change Test Settings a Select SET PARAMETERS u...

Страница 21: ...the keys d Press the key to save changes to the Bypass setting For more details about Bypass Settings see page 36 Prior to commencing patient testing remove the Supervisor CODE CHIP module from the me...

Страница 22: ...e information is transferred into the Quidel Triage MeterPro s memory A CODE CHIP module typically needs to be installed only once and remains in the meter memory A CODE CHIP module does not need to r...

Страница 23: ...or Disable Auxiliary ID Number of Characters in Misc Test ID DISPLAY SETTINGS Language Printer Mode Auto Power Off Display Contrast Printer Contrast COMMUNICATIONS Baud Rate Enable or Disable LIS Enab...

Страница 24: ...see a list of programmable parameters METER SETTINGS CLOCK USER ID RANGES TEST SETTINGS QC PARAMETERS BYPASS SELECT WITH v OR AND PRESS ENTER OR PRESS EXIT The Set Parameters Menu 4 Use the keys to se...

Страница 25: ...X column 8 Use the keys to choose the largest number of characters that will appear in the User ID The maximum number of characters possible is 16 9 Press the key to save changes Characters Pat ID Thi...

Страница 26: ...aracters that will appear in the Auxiliary ID The maximum number of characters possible is 12 9 Press the key to save changes Aux ID Enable Disable This setting will turn on or turn off the Auxiliary...

Страница 27: ...rgest number of characters that will appear in the MISC TEST ID The maximum number of characters possible is 20 9 Press the key to save changes Display Settings Language This setting will determine wh...

Страница 28: ...Power Off 6 Use the keys to choose the desired timer setting for the meter to power itself off when not in use The choices are 1 2 HOUR 1 HOUR 2 HOURS 4 HOURS or NONE If NONE is selected the meter mus...

Страница 29: ...n entered will allow data to be transmitted Baud Rate 1 1 1 1 From a list of programmable parameters choose METER SETTINGS using the keys 2 Press the key 3 Use the keys to select COMM SETTINGS 4 Press...

Страница 30: ...S Enable 1 From a list of programmable parameters choose METER SETTINGS using the keys 2 Press the key 3 Use the keys to select COMM SETTINGS 4 Press the key 5 Use the keys to select LIS 6 Use the key...

Страница 31: ...o the DATE 9 Using the keypad type in the 6 digit date according to the DATE FORMAT below it 10 Press to move to the DATE FORMAT 11 Use the keys to select MM DD YY DD MM YY or YY MM DD The previously...

Страница 32: ...er ID you want to update To correct press to clear the entire ID or select the incorrect character by using the keys Then type in the correct ID 6 Press the key 7 The user ID is displayed with the exp...

Страница 33: ...ing the month number the day number and the year number in the format you have chosen The numbers you type will appear from the right and move left as you continue to type To correct press to clear th...

Страница 34: ...upervisor may wish to control which tests are available to be run at the time of patient testing The options are 1 ACTIVE ACTIV ON SCREEN The test is always run and cannot be deselected 2 INACTIVE INA...

Страница 35: ...tive tests is 2 or 3 standard deviations QC Device Frequency The QC DEVICE FREQUENCY setting determines the maximum interval between QC Device tests that a user is allowed to run patient tests When th...

Страница 36: ...ts 1 From the list of programmable parameters choose QC PARAMETERS using the keys 2 Press the key 3 Use the keys to select the information to the right of NUM QUANT STD DEVS 4 Using the keys select on...

Страница 37: ...mple CODE CHIP module included in each box of controls Lot number begins with a C QC Device CODE CHIP module included in the black QC Device box comes with the Quidel Triage MeterPro and is labeled wi...

Страница 38: ...rPro s memory 6 Press to return to acknowledge the message 7 Remove the CODE CHIP module from the CODE CHIP module Port 8 Place the CODE CHIP module back into its original container for storage INSTAL...

Страница 39: ...software upgrade Supervisor Installation Procedure Otherwise No restriction SET PARAMETERS Access Control Supervisor Programmable Settings Supervisor DELETE RESULTS Patient Test Results Supervisor Unl...

Страница 40: ...day of patient testing To run the QC Device test 1 Press the key to power the Quidel Triage MeterPro on 2 Select RUN TEST using the keys 3 Press the key 4 Type in your identification User ID To corre...

Страница 41: ...e a printed copy of the results 13 Remove the QC Device from the Meter and place in the QC Device Box DO NOT DISCARD THE QC DEVICE IMPORTANT Important instructions for all tests a Gently insert the te...

Страница 42: ...idel QC Sample To run the Quality Control sample 1 Press the key to power the Quidel Triage MeterPro on 2 Select RUN TEST using the keys 8 23AM 01 20 18 RUN TEST RECALL RESULTS INSTALL NEW CODE CHIP S...

Страница 43: ...mpt to pull it out The device may be ejected by returning to the Main Menu and pressing EXIT 12 Repeat steps for each quality control sample IMPORTANT Important instructions for QC Sample Tests D If t...

Страница 44: ...correct press to clear the entire ID or select the incorrect character by using the keys Then type in the correct ID 5 Press the key 6 Select PATIENT SAMPLE using the keys PATIENT SAMPLE QC DEVICE QC...

Страница 45: ...st device has been pulled into the meter the meter will display a message similar to the following SELECT TESTS CARDIAC CKMB 2 MYO 3 TNI PRESS TO PAUSE OR 0 9 FOR TEST PRESS ENTER TO CONTINUE OR EXIT...

Страница 46: ...he Main Menu and pressing EXIT IMPORTANT Important instructions for patient tests a If the test device s Internal Quality Control QC results are unacceptable the results for the affected analyte s wil...

Страница 47: ...TEST SELECT WITH v OR AND PRESS ENTER PRESS EXIT TO EJECT DEVICE 7 Press the key 8 Type or scan the Misc Test identification Misc Test ID To correct press to clear the entire ID or select the incorre...

Страница 48: ...has been inserted properly b Press the key to start the test c The Quidel Triage MeterPro pulls in the test device and scans it d When the test is complete the meter will beep and display the results...

Страница 49: ...talled to view the full User ID Last Record Purpose Last Record is used to retrieve the test results for the last test run To retrieve the last record 1 Press the key to power the Quidel Triage MeterP...

Страница 50: ...the selected panel will be displayed For individual patient results a Select Patient ID using the keys b Type or scan in the patient ID To correct press the key to clear the entire ID or select the i...

Страница 51: ...to select the desired panel 10 Select DEVICE L N using the keys 11 To display all results for the selected panel press the key 12 To display results specific to one lot type in the lot number of the t...

Страница 52: ...press the key 12 To display results specific to one User ID type in the User ID To correct press to clear the entire ID or select the incorrect character by using the keys Then type in the correct ID...

Страница 53: ...e selected panel are displayed 15 Use the keys to view the entire record 16 Press the key to make a printed copy 17 Press the key to upload the selected results to LIS Till Date To retrieve results fr...

Страница 54: ...s of these quality control QC tests can be retrieved from the Quidel Triage MeterPro s memory Quality Control QC test Results can be retrieved by QC Sample Results Panel Type QC Sample Lot Number L N...

Страница 55: ...press to clear the entire number Then type in the correct number 12 Press the key 13 All results from test run on the requested QC Sample lot number for the selected panel are displayed 14 Use the key...

Страница 56: ...ve QC results from tests run on the Quidel Triage MeterPro starting with the user you select 1 Press the key to power the Quidel Triage MeterPro on From the MAIN MENU select RECALL RESULTS using the k...

Страница 57: ...he entire number Then type in the correct number 12 Press the key 13 All results from tests run from the date selected for the selected panel if QC Sample are displayed 14 Use the keys to view the ent...

Страница 58: ...ackground If the QC for the lot has failed the word FAILED will appear in place of an expiration date If a CODE CHIP has been installed for a new lot but QC has yet to be established the words NOT RUN...

Страница 59: ...Select MISC TEST ID using the keys b Type or scan in the Misc Test ID To correct press the key to clear the entire ID or select the incorrect character by using the keys Then type in the correct ID c...

Страница 60: ...the keys 9 To display all results for the selected panel press the key 10 To display results specific to one lot type in the lot number of the test device To correct press the key to clear the entire...

Страница 61: ...incorrect character by using the keys Then type in the correct ID 11 Press the key 12 All results from tests run by the requested user for the selected panel are displayed 13 Use the keys to view the...

Страница 62: ...power the Quidel Triage MeterPro on 2 From the MAIN MENU select RECALL RESULTS using the keys 3 Press the key 4 Select MISC TEST RESULTS using the keys 5 Press the key 6 Select PANEL TYPE using the k...

Страница 63: ...to clear the entire ID or select the incorrect character by using the keys Then type in the correct ID 9 Press the key to make a printed copy QC Sample Results To print all QC Sample Results 1 Press...

Страница 64: ...SULTS using the keys 5 Press the key 6 Select MISC TEST RESULTS using the keys 7 Press the key to make a printed copy Upload Results to LIS When interfaced directly to an LIS selecting this feature wi...

Страница 65: ...XXX MORE RECORDS PRESS ENTER The Memory Alert Screen IMPORTANT The Delete function permanently removes all patient data from the meter memory Before continuing you may wish to first print results or u...

Страница 66: ...want to delete all patient results press the key Display Backlight The Quidel Triage MeterPro has the capability for the operator to backlight the display for providing improved clarity depending on a...

Страница 67: ...o alpha character is displayed 6 M N O The first press of a button causes the first letter to be displayed The second press of the same button within three seconds causes the second letter to be displ...

Страница 68: ...er at an angle as the printer must sense the edge of the paper along the feed path 5 Insert the new roll of paper into the paper compartment 6 Position the paper such that the paper will feed from und...

Страница 69: ...Position the paper such that the paper will feed from under the roll as opposed to over the top of the roll see picture above 7 Insert the paper edge under the paper roller platen until it firmly seat...

Страница 70: ...th front and back of the door Check Battery Level The Quidel Triage MeterPro checks the active power source after ejecting a device or printing To check the battery power navigate to the MAIN MENU dis...

Страница 71: ...which analytical methods are traditionally controlled is through the use of liquid control specimens containing known amounts of analytes Over the last 15 years advances in technology have changed th...

Страница 72: ...ontrol features provides more information about the validity of each assay than can be provided by the more traditional methods of reagent and assay procedure validation Quality Assurance Functions 1...

Страница 73: ...izontal and vertical alignment of the zone measurement to verify laser alignment and device transport The QC Device also confirms laser stability A Bar Code is printed on the bottom of every test devi...

Страница 74: ...r Total Quality Assurance features inherent in the Quidel Triage tests The Quidel Triage tests have been designed to maximize Total Quality Assurance in any testing environment The combination of the...

Страница 75: ...the RECALL RESULTS menu 5 Supervisor Access To prevent misuse the Supervisor CODE CHIP module is required for access to the SET PARAMETERS and DELETE RESULTS functions of the software The Supervisor C...

Страница 76: ...iode Memory Capacity 750 Patient Records 32 Reagent Lot CODE CHIP modules 200 QC Sample Results 200 QC Sample CODE CHIP modules 70 QC Device Results 4 QC Device CODE CHIP modules 250 Misc Test Results...

Страница 77: ...ced Contact Quidel CODE CHIP module will not fit in meter a CODE CHIP module may have been inserted upside down a Re insert CODE CHIP module right side up b CODE CHIP module or Meter contacts could be...

Страница 78: ...e eliminates the message the original CODE CHIP is at fault discard CODE CHIP module b Meter error b If the new CODE CHIP module does not work install a known working CODE CHIP module to verify proper...

Страница 79: ...ed when the meter has been without power Press the ENTER key QC Device calibration alignment or laser failed Messages Calib Pass or Fail Laser Pass or Fail Align Pass or Fail a May be due to lint or d...

Страница 80: ...nd Electronic Equipment WEEE Appendix D Glossary Analyte Chemical substance being measured Assay Test Calibration Comparison of result to a reference standard Character One part of a longer string of...

Страница 81: ...Device Not all Quidel Triage test panels have the MMX feature Parameters Options functions Photodiode Detects the light emitted from the fluorescent dye in the test devices used in the Quidel Triage...

Страница 82: ...DE CHIP module used to download new software onto the Quidel Triage MeterPro Supervisor CODE CHIP module The CODE CHIP module that allows a supervisor to set parameters and delete results Test device...

Страница 83: ...Page 83 of 86...

Страница 84: ...Page 84 of 86...

Страница 85: ...Page 85 of 86...

Страница 86: ...l Triage MeterPro MDSS GmBH Schiffgraben 41 30175 Hannover Germany Quidel Cardiovascular Inc 9975 Summers Ridge Drive San Diego CA 92121 USA quidel com ENSRC26611enA PN 26611en Rev A 2018 03 Revision...

Отзывы: