Page
44
of
86
d.
If
an
individual
analyte
is
outside
the
specified
range,
the
Quidel
Triage
MeterPro
allows
the
user
to
rerun
only
the
failed
analyte
on
the
next
QC
Sample
Test.
When
all
analytes
have
passed,
the
QC
Sample
timer
is
set
to
the
date
of
the
first
passing
analyte.
Patient
Sample
1.
Press
the
key
to
power
the
Quidel
Triage
MeterPro
on.
2.
Select
RUN
TEST
using
the
keys.
8:23AM
01•20•18
RUN
TEST
RECALL
RESULTS
INSTALL
NEW
CODE
CHIP
SELECT
WITH
v
OR
^
AND
PRESS
ENTER
PRESS
EXIT
TO
EJECT
DEVICE
3.
Press
the
key.
4.
Type
or
scan
in
your
identification
(User
ID).
To
correct,
press
to
clear
the
entire
ID,
or
select
the
incorrect
character
by
using
the
keys.
Then
type
in
the
correct
ID.
5.
Press
the
key.
6.
Select
PATIENT
SAMPLE
using
the
keys.
PATIENT
SAMPLE
QC
DEVICE
QC
SAMPLE
MISC.
TEST
SELECT
WITH
v
OR
^
AND
PRESS
ENTER
OR
PRESS
EXIT TO
EJECT
DEVICE
7.
Press
the
key.
8.
Type
or
scan
the
patient’s
identification
(Patient
ID).
To
correct,
press
to
clear
the
entire
ID,
or
select
the
incorrect
character
using
the
keys.
Then
type
in
the
correct
ID.
9.
Press
the
key.
Main
Menu
Run
Test
NOTE
:
If
the
User
ID
bypass
is
ON
,
the
meter
will
not
ask
you
to
enter
a
User
ID.
NOTE
:
If
the
Bar
Code
Scanner
is
being
used,
simply
point
the
scanner
at
the
bar
code,
press
the
button
on
the
scanner
handle
and
wait
for
the
beep.
The
ID
will
appear
on
the
display