IMPORTANT:
Important
Instructions
for
QC
Device
tests:
Dust,
lint,
fibers
and
other
small
particles
may
interfere
with
the
QC
Device.
Keep
the
QC
Device
free
of
contaminants.
The
QC
Device
is
light
Ͳ
sensitive
and
should
be
stored
in
its
black
opaque
case
when
not
in
use.
If
the
QC
Device
tests
fail,
wipe
the
QC
Device
clean
with
a
lint
free
cloth
to
remove
any
oils,
dust,
fibers,
or
fingerprints.
Do
not
apply
any
liquid
to
the
QC
Device.
After
cleaning
the
device,
repeat
the
QC
Device
test.
If
the
QC
Device
test
fails
after
you
have
cleaned
the
device,
contact
Quidel.
QC
Sample
To
run
the
Quality
Control
sample:
1.
Press
the
key
to
power
the
Quidel
Triage
MeterPro
on.
2.
Select
RUN
TEST
using
the
keys.
8:23AM
01•20•18
RUN
TEST
RECALL
RESULTS
INSTALL
NEW
CODE
CHIP
SELECT
WITH
v
OR
^
AND
PRESS
ENTER
PRESS
EXIT
TO
EJECT
DEVICE
The
Main
Menu
3.
Press
the
key.
4.
Type
in
your
identification
(User
ID).
To
correct,
press
the
key
to
clear
the
entire
ID,
or
select
the
incorrect
character
by
using
the
keys.
Then
type
in
the
correct
ID.
5.
Press
the
key.
6.
Select
QC
Sample
using
the
keys.
PATIENT
SAMPLE
QC
DEVICE
QC
SAMPLE
MISC.
TEST
SELECT
WITH
v
OR
^
AND
PRESS
ENTER
OR
PRESS
EXIT TO
EJECT
DEVICE
Menu
screen
as
it
looks
when
operator
selects
QC
Sample.
NOTE:
If
the
User
ID
bypass
is
ON
,
the
meter
will
not
ask
you
to
enter
a
User
ID.
Main
Menu