Contrast
This
setting
will
adjust
the
printer
contrast
for
meters
with
a
serial
number
29901
or
greater.
1.
From
a
list
of
programmable
parameters,
choose
METER
SETTINGS
using
the
keys.
2.
Press
the
key.
3.
Use
the
keys
to
select
PRINTER
CONTRAST
.
4.
Press
the
key.
5.
Use
the
keys
to
select
the
number
to
the
right
of
Contrast.
6.
Use
the
keys
to
change
the
contrast
of
the
meter’s
screen.
The
range
is
from
0
Ͳ
8
(8
being
the
greatest
contrast).
7.
Press
the
key
to
save
changes.
Communication
Settings
Prior
to
activating
the
LIS
feature,
obtain
the
meter’s
serial
number
located
underneath
the
meter
and
contact
Quidel
(Refer
to
the
Assistance
section).
The
Quidel
Representative
will
provide
you
with
a
password
that,
when
entered,
will
allow
data
to
be
transmitted.
Baud
Rate
1.
1.
1.
1.
From
a
list
of
programmable
parameters,
choose
METER
SETTINGS
using
the
keys.
2.
Press
the
key.
3.
Use
the
keys
to
select
COMM
SETTINGS.
4.
Press
the
key.
5.
Use
the
keys
to
select
BAUD
RATE.
6.
Use
the
keys
to
choose
9600
or
38400
.
7.
Press
the
key.
Patient
Result
Approval
Some
institutions
may
wish
for
patient
test
results
to
be
verified
by
the
user
before
printing
or
sending
to
the
LIS.
Activation
of
this
feature
displays
the
result,
then
requires
the
user
to
accept
or
reject
the
result
before
the
record
is
printed
and
saved.
If
the
test
is
rejected,
the
meter
will
permanently
flag
it
as
rejected
by
the
user.
1.
From
a
list
of
programmable
parameters,
choose
METER
SETTINGS
using
the
keys.
2.
Press
the
key.
Main
Menu
Set
Parameters
Meter
Settings
Comm
Settings
Main
Menu
Set
Parameters
Meter
Settings
Display
Settings
Main
Menu
Set
Parameters
Meter
Settings
Comm
Settings
Содержание Triage MeterPro
Страница 1: ...Page 1 of 86 User Manual...
Страница 13: ...Page 13 of 86 Parts of the Quidel Triage MeterPro...
Страница 83: ...Page 83 of 86...
Страница 84: ...Page 84 of 86...
Страница 85: ...Page 85 of 86...