Introduction
3
Roll the level toward you so that the display is facing you, but the lettering on the face of the unit is
upside down. Be sure to set the level in the exact spot as in Step 1. Wait 10 seconds for the level to
completely settle and take note of the angle on the display.
4
Rotate the level end-for-end so the display is facing away from you. The screw on the back of the level
should be on the right side. Be sure to set the level in the exact spot as in Step 1. Wait 10 seconds for
the level to completely settle and take note of the angle on the display.
5
Compare the level display readings that you captured from Steps 1 through 4. If any of the four readings
vary from one another more than 0.1 degree, you must re-calibrate the level. For more information, see
"Calibrating the Digital Level" in the
Scalar i6000 Installation Guide
. If the variance between the
readings are within the 0.1 degree limit, the level is within compliance and is ready to be used.
Verifying Level Condition
Required tools: digital level
1
In the control module only, verify the system unit is level front to back and left to right by placing the
digital level at the following locations:
a. Inside the access door
b. On the left side inside the access door
A reading on the digital level that is 0.00 +/- 0.30 is to be considered level.
Содержание Scalar i6000
Страница 14: ...14 Introduction Figure 1 Power Rail Position in Module Two bridge clamps connect adjacent rails...
Страница 15: ...Scalar i6000 Installation Guide Super Doc 15 Figure 2 Power Rail Front 4 pin and 3 pin connectors...
Страница 23: ...Scalar i6000 Installation Guide Super Doc 23 Figure 6 Installing Bottom and Middle Racks...
Страница 24: ...24 Introduction Figure 7 Middle X Axis Gear Rack...