Internal jumpers and switches
See figure 17 for the location of jumpers and switches on the cell controller board.
CAUTION
Internal jumpers and switches should not be altered without first consulting your supplier or
Pyramid Technical Consultants, Inc. Unpredictable operation may result.
13.1 Jumpers
Jumper bank 2 informs the A500 about various hardware options that may be installed. These
jumpers are normally factory-configured and should not be changed, and depend upon the
hardware revision of the A500 board.
A500 Hardware Rev 1
Link
Description when fitted
0-2
The type of application specific I/O board, if any, that is connected to I/O port #1.
Connect these links to form a binary number from 0-3.
0 [link none connected]
-no I/O board present
1 [link 0 connected]
-GC board connected
2 [link 1 connected]
-G board connected
3-4
The type of fiber-optic mezzanine board that is installed. Connect these links to
form a binary number from 0-1.
0 [none connected]
-no mezzanine board present
1 [link 3 connected]
-RX10 board present
2 [link 4 connected]
-R5T5 board present
3 [link 3-4 connected]
-TX10 board present
A500 Hardware Rev 2 and Rev 3
Link
Description when fitted
0-2
The type of application specific I/O board, if any, that is connected to I/O port #1
(channel B). Connect these links to form a binary number from 0-3. The currently
supported boards are:
0 [none connected]
-no I/O board present
1 [link 0 / B3 connected]
-GC board connected
2 [link 1 / B2 connected]
-G board connected
3-5
The type of application specific I/O board, if any, that is connected to I/O port #2