Reset The Lamp Life Hour Counter
When the life of the UV lamp is exhausted and is replaced, the controller needs to reset the
operating hours of the lamp by performing the following procedure.
1. Ensure that the replace lamp error message is being
displayed and that the alarm has been acknowledged
by following the instructions in the previous section.
2. At the same time, press and hold the Left and Right
keys for 6 seconds.
The system will ask for confirmation of reset.
3. Press the Enter key. System will BEEP as an audio
confirmation that the process has been completed
successfully. Display will now show 0, indicating that
the hour counter has been reset.