14
Installation & Operation Manual
Copyright © Puretec Pty Ltd 2022
5. While holding the guide rod in place at the opposite
end, carefully slide the sleeve out of chamber. If it
feels tight, gently rotate the quartz sleeve while also
pulling it out.
Make sure that the sleeve is not at an angle as
otherwise pressure will be applied on the sides of the
sleeve and against the reactor chamber causing the
sleeve to fracture.
WARNING: USE THE GUIDE ROD AS A SUPPORT OR ELSE THE QUARTZ SLEEVE
CAN “FALL” INSIDE THE REACTOR WHILE REMOVING AND BREAK.
6. Once the quartz sleeve is removed from the system it can either be cleaned or replaced
with a new sleeve as required.
Cleaning The Quartz Sleeve
The quartz sleeve forms the barrier between the UV lamp and the water under treatment. It
is essential that the sleeve is kept clean otherwise there will be a reduction of the UV light
which will reduce the effectiveness of treatment. We recommend using Puretec SC1000 UV
Sleeve Cleaner to clean quartz sleeves.
Steps
1. The quartz sleeve should not be handled without gloves. Only use powder-free gloves.
2. If it is lightly soiled spray the sleeve with SC1000 and wipe it down with a lint-free cloth.
3. If the quartz sleeve is heavily soiled soak it in SC1000 for 30 minutes and then wipe
along the length of the quartz sleeve with a lint-free cloth until clean.
4. Inspect the interior of the sleeve to make sure it is clean and dry.
5. Install/reinstall the cleaned or new quartz sleeve in reverse order.
DO NOT FORGET TO INSTALL THE QUARTZ SEALING O-RING ONTO THE SLEEVE
UNTIL IT RESTS AGAINST THE END PLATE BEFORE RETIGHTENING THE SEALING NUT.
PURETEC RECOMMENDS INSTALLING NEW QUARTZ SEALING O-RINGS EVERY TIME
THE QUARTZ IS REMOVED OR REPLACED.
UV Intensity Sensor (Optional)
128
95
Maintenance