9.3 Testing High Pressure Alarm
NOTE:
All testing values are based on default Dehydrator settings, if settings
have been changed, adjust testing values accordingly.
9.3.1
Make note of the current
High Pressure setting.
(Section 8.4.5.2 )
9.3.2
Set High Pressure (Section
8.7.3 )
over 20.7 KPa.
After one (1) minute, the
High Pressure Alarm should
appear on the display.
9.3.3
Set High Pressure back to
setting recorded in step 9.3.1
9.3.4
Press the
RESET
Button
to
clear the alarm.
If you are unable to create a High Pressure Alarm as described, see Section 11.6
for troubleshooting information.