DIAMENTOR E2 

Cleaning, Preventive Maintenance and Disposal 

 

D651.131.00/05 en

 

23

 

9  Cleaning, Preventive Maintenance and Disposal 

9.1 Cleaning  

Ionization chambers 

The outer shells of the ionization chambers are 
made of macrolon (polycarbonate). The electrode 
surfaces are made conductive by vacuum metalliza-
tion. In this process a transparent metal coating is 
applied to the electrode. In order to keep the coat-
ing transparent, ionization chambers must not be 
treated or come into contact with alcohol or sol-
vents.  

Cleaning agents and solvents contained in disinfec-
tants (alcohols) have proved to be particularly 
harmful.  

 

Wipe the chambers clean with a moist cloth. 

 

Please use a common disinfection agent for 
medical instruments that is listed in your national 
hygiene institute (e.g., VAH in Germany). If your 
country does not provide such a list we recom-
mend using either an aldehyde based solution or 
a solution with a quaternary ammonium com-
pound. 

Surface 

CAUTION 

Equipment damage -

 Do not use disinfectants 

on a phenol base or peroxide compounds to 
disinfect the surface. 

 

 

First of all, disconnect the devices from the 

power line, before attempting to clean or disin-
fect them. 

 

Rub the devices down with a moist cloth; no 

liquid must enter the devices. Due to the sur-
face geometry, a thorough disinfection is not 
possible. The device cannot be sterilized. 

Connection cables 

 

disconnect the cables from the devices, pulling 

on the plug, not on the cable 

 

for cleaning, rub the cables down with a cloth 

moistened with soap water or disinfectant 

 

never immerse the cable in liquid! 

9.2 Preventive Maintenance 

WARNING

 

Hazards to Persons! Equipment Damage! 

Only specially trained staff is allowed to perform 
the following service tasks. 

Further service tasks may only be carried out by 
PTW-Freiburg or by persons authorized by 
PTW-Freiburg. 

 

Check before each use 

  Before each use, visually inspect the device 

and the cables for signs of mechanical dam-
age 

  Also, conduct a constancy check as described 

in section 5.1. 

  If damages or malfunctions are identified, the 

device must be repaired before it is used 
again. 

Repair  

Repairs may only be carried out by PTW-Freiburg 
or by persons / companies authorized by 
PTW-Freiburg.  

Содержание DIAMENTOR E2

Страница 1: ...User Manual DIAMENTOR E2 T11033 from SN 02000 and higher D651 131 00 05 en 2012 11 G Hn Kau...

Страница 2: ...se use has been ex pressly permitted by PTW Freiburg Safe op eration and proper product performance are not guaranteed if accessories or consumables from other manufacturers are used PTW Freiburg cann...

Страница 3: ...ance with a quality management system which is continu ously updated according to national and inter national standards PTW FREIBURG Physikalisch Technische Werkst tten Dr Pychlau GmbH L rracher Str 7...

Страница 4: ...16 5 PERFORMING A MEASUREMENT 17 5 1 Constancy Check 17 5 2 Measurement 17 5 3 Air Density Correction 18 5 4 Print out 18 6 ERROR MESSAGES AND TROUBLESHOOTING 19 7 CALIBRATION 20 8 PERIPHERALS 22 8 1...

Страница 5: ...to connect a label printer or a PC This manual only describes the DIAMENTOR E2 A detailed description of the RS232 interface is supplied on request document number D651 131 1 2 For Your Safety The fol...

Страница 6: ...evice from the power line first remove the connector from the wall outlet Then disconnect the cable from the device Devices on which moisture condensation has developed as a result of temperature chan...

Страница 7: ...1 ANSI AAMI ES 60601 1 CAN CSA C22 2 No 60601 1 WARNING Suffocation Hazard Dispose of the packag ing material observing the applicable waste control regulations Keep the packaging mate rial out of chi...

Страница 8: ...d to the power line or the cable is clamped into the DC supply the system is powered even if the DIAMENTOR E2 is turned off CAUTION Equipment damage Handle the product with care to avoid equipment dam...

Страница 9: ...n Association for Electrical Electronic and Information Tech nologies The product is certified by UL Un derwriters Laboratories for both the U S and Canadian markets to the applicable U S and Canadian...

Страница 10: ...lation Inside the patient Outside the patient Not medical used area environment environment Figure 2 Note for installation in the patient environment PTW DIAMENTOR E2 T11033 Printer PC IEC 60950 1 Pri...

Страница 11: ...em A 4 Connection adapter 5 On Off switch Device switched on Device switched off 6 Display 8 digit seven segment display 7 TEST button to conduct a device test and for cursor control in the menu mode...

Страница 12: ...the supply voltage WARNING The system has no means of isolation from the supply voltage When the mains plug is con nected to the power line or the cable is clamped into the DC supply the system is po...

Страница 13: ...xternal supply voltage CAUTION Make sure that the release of the installation manufacturer is on your hand before you con nect the DIAMENTOR E2 via power limiter T26060 to the installation WARNING The...

Страница 14: ...orities where the equipment is sold The complete system has to be evaluated ac cording to the requirements of IEC 60601 1 ANSI AAMI ES 60601 1 CAN CSA C22 2 No 60601 1 by the end user To achieve the r...

Страница 15: ...itch refer to Figure 3 item 5 The firmware version will be displayed for about 2 seconds followed by a display test After completion of the test the DIAMENTOR E2 is operational 4 4 Mounting the ioniza...

Страница 16: ...amber you should enter the new chamber calibration factor as described below To change the chamber calibration factor confirm the setting for the measuring quantity with RESET The menu item for settin...

Страница 17: ...rsion has been displayed for about 2 seconds and the display test has been completed success fully measured values are displayed according to the selected mode as soon as radiation is detected Auto me...

Страница 18: ...T p T p KD p0 1 013 hPa p current air pressure T0 293 K 20 C T current temperature According to device the correction can perhaps be executed over a correction menu or an used soft ware 5 4 Print out...

Страница 19: ...cation error Turn the device off and on again When the message appears again the device must be repaired Error 54 Memory malfunction Turn the device off and on again When the message appears again the...

Страница 20: ...ions where there is no absorber between chamber and patient is up to 9 In conjunction with undercouch tubes the absorption of the patient couch situated between the patient and the chamber is equivale...

Страница 21: ...e Repeat the measurement 3 times noting down the results Fluoroscopy overcouch tube Place the measurement stand without detector on the patient couch phantom at the top Start the fluoroscopy and deter...

Страница 22: ...IAMENTOR E2 and the measured value 8 2 PC WARNING Please read the printer manual before putting the printer into operation Operate the DIAMENTOR E2 in the patient en vironment only if the connected PC...

Страница 23: ...ctants on a phenol base or peroxide compounds to disinfect the surface First of all disconnect the devices from the power line before attempting to clean or disin fect them Rub the devices down with a...

Страница 24: ...iburg or a qualified calibration laboratory Inspections must also be performed after repairs or after each influence which may have changed the behavior of the product Calibration The exact calibratio...

Страница 25: ...1 Measuring principle detection of irradiation by means of ionization chambers analog and digital signal processing digital readout on LED Operating mode continuous operation Measuring quantities Dos...

Страница 26: ...3 mm x 36 mm x 151 mm Weight 250 g Adapter Power supply T99052 Input 100 240 VDC 0 33 0 18 A 47 63 Hz Output 24 VDC 0 62 A Power limiter T26060 Input 24 VDC 10 Output 24 VDC 10 9 VA Environmental cond...

Страница 27: ...for cursor control to change and confirm settings and to quit a menu Electrical calibration When the Service menu is called up the current setting for the electrical calibration factor CAL will be sho...

Страница 28: ...ic environment specified below The customer or the user of the DIAMENTOR E2 should assure that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance Radio Fr...

Страница 29: ...1000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commer cial or hospital environment Voltage dips short in...

Страница 30: ...meters m Field strengths from fixed RF transmitters as deter mined by an electromagnetic site survey a are less than the compliance level in each frequency range b Interference may occur in the vicin...

Страница 31: ...out put power of transmit ter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d P V1 5 3 1 2 P 80 MHz to 800 MHz d P E1 5 3 1 2 P 800 MHz to 2 5 GHz d P E1 7 2 3 P 0 0...

Страница 32: ...he toxic or hazardous substances or elements contained in electronic information products will not leak or mutate under normal operating conditions so that the use of such electronic information produ...

Страница 33: ...s below the limit requirement in SJ T11363 2006 X Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirem...

Страница 34: ...per phase 10 IEC 61000 3 3 Electromagnetic compatibility EMC Part 3 3 Limits Limitation of voltage changes voltage fluctuations and flicker in public low voltage supply systems for equipment with rate...

Страница 35: ...28 Error Messages 19 F First Time Operation 12 G General Information 2 I Intended Use 5 Ionization chamber mounting 15 L Literature 34 M Maintenance 23 Measurement 17 O Operating Controls and Indicato...

Отзывы: