DIAMENTOR E2
First Time Operation
14
D651.131.00/05 en
Installation of the power limiter T26060 (refer to
Figure
5
)
The DIAMENTOR E2 may be operated only with
power limiter T26060 in compliance with the follow-
ing conditions.
•
The earth and enclosure leakage currents of the
whole installation have to be measured when
putting into operation. The earth and enclosure
leakage current may not exceed the maximum
values of 300 µA (safety class I) or 150 µA
(safety class II) depending on the safety class.
•
The user replaceable mains (line) fuse must be
an approved type acceptable to the authorities
where the equipment is sold.
•
The complete system has to be evaluated ac-
cording to the requirements of IEC 60601-1
ANSI/AAMI ES 60601-1/
CAN/CSA-C22.2.No.60601-1 by the end user.
•
To achieve the requirements for electro-
magnetic compatibility it is recommended to
connect the DC power supply output to GND
and earth (refer to Figure 5). This earth connec-
tion
has no protective function (function earth).
Maximum Earth and Enclosure Leakage Current
Measurement according to IEC 62353
Class I equipment:
The maximum leakage current is 300 µA.
Class II equipment:
The maximum leakage current is 150 µA.
Figure 5: Installation of the power limiter T26060
T26060
PTW
brown
white
24 VDC
GND
Earth Connection
2 MOPP
24 VDC ±10%
Double / Reinforced
insulation
IEC 60601-1
Power Supply
Power GND