DIAMENTOR E2
Electromagnetic Compatibility (EMC) according to IEC 60601-1-2
D651.131.00/05 en
29
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The DIAMENTOR E2 is intended for use in the electromagnetic environment specified below. The customer or
the user of the DIAMENTOR E2 should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environ-
ment - Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV
air
± 6 kV
contact
± 8 kV
air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humid-
ity should be at least 30 %.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV
for power supply lines
± 1 kV
for input / output lines
± 2 kV
for power supply lines
± 1 kV
for input / output lines
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Surges
IEC 61000-4-5
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Voltage dips, short inter-
ruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
(
tested using power supply
T99042
)
< 5 % U
T
(> 95 % dip in U
T
) for
0.5 cycle
40 % U
T
(60 % dip in U
T
) for
5 cycles
70 % U
T
(30 % dip in U
T
) for
25 cycles
< 5 % U
T
(> 95 % dip in U
T
) for 5 s
< 5 % U
T
(> 95 % dip in U
T
) for
0.5 cycle
40 % U
T
(60 % dip in U
T
) for
5 cycles
70 % U
T
(30 % dip in U
T
) for
25 cycles
< 5 % U
T
(> 95 % dip in U
T
) for 5 s
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the
DIAMENTOR E2 requires
continued operation during
power mains interruptions, it
is recommended that the
DIAMENTOR E2 be pow-
ered from an uninterruptible
power supply or a battery
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical com-
mercial or hospital environ-
ment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.